Skip to the content

FDA Medical Device Recalls (openFDA)

Draeger, Inc.: Brand Name: Anesthesia Circuit Kit Product Name: Anesthesia Circuit Kit Flex 6 Model/Catalog Number: MP02744 Software Version: Not applicable Product Description: Accessory for conducting medical gases between anesthesia machine and patient for anesthetic use. Kit is intended for single use only

Recall number
Z-1497-2025
Recalling firm
Draeger, Inc.
Product
Brand Name: Anesthesia Circuit Kit Product Name: Anesthesia Circuit Kit Flex 6 Model/Catalog Number: MP02744 Software Version: Not applicable Product Description: Accessory for conducting medical gases between anesthesia machine and patient for anesthetic use. Kit is intended for single use only
Status
Open, Classified
Initiated
2025-03-12
Posted
2025-04-03
Terminated
not on file
Reason
The potential for cracks forming in the breathing circuit hose.
Root cause
Under Investigation by firm
Action
On March 12, 2025 URGENT MEDICAL DEVICE RECALL letters were sent to customers. Actions to be Taken When using these breathing circuits: 1. Use holders with a larger contact surface (see Figure 2 above). 2. Perform leak testing after stretching the flexible hose to the required length for the application. Please observe the products Instructions for Use (IFU) According to our records, you have received at least one of the listed breathing circuits. Please inform all potential users in your facility. If you have forwarded affected products to third parties, please also forward this Urgent Medical Device Recall notification. Please also complete the enclosed Medical Device Recall Return Response Acknowledgment and Receipt Form to confirm to us that you have received this information. Adverse events or quality problems experienced with the use of this product may be reported to FDA s MedWatch Adverse Event Reporting program either online or by phone at 1-800-FDA-1088. We apologize for any inconveniences resulting from this action. If you have any questions regarding this letter, please contact Michael Kelhart between the hours of 8:00 AM 4:30 PM EST at 267-664-1131 or via email at mike.kelhart@draeger.com. The US Food and Drug Administration has been advised of this action. ****Update 07/08/2025****On June 25, 2025 URGENT MEDICAL DEVICE RECALL - UPDATE letters were sent to consignees to clarify and provide complete UDI information for the device to include both primary and secondary UDI numbers as errors were identified in the initial letter.
Quantity
293660 units
Distribution
Domestic: Nationwide Distribution. International - following countries: Austria, Belgium, Bulgaria, China, Croatia, Czech Republic, France, Germany, Ghana, Greece, Guatemala, Hungary, India, Italy, Japan, Panama, Philippines, Portugal, Singapore, South Africa, Spain, Sweden, United Kingdom
Lot, serial or model codes
Model/Catalog Number: MP02744; UDI Number 040486775389675; All lot numbers.
510(k) numbers
not on file
PMA numbers
not on file
Product code
OFP
Firm FEI
2517967
Firm city
Telford
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register