Skip to the content

FDA Medical Device Recalls (openFDA)

Merit Medical Systems, Inc.: 7F Prelude Short Sheath Introducer. Catalog Numbers: K15-00070, K15-00170, PSS-7F-4-035MT, PSS-7F-4MT. Produc Usage: The Merit Prelude Short Sheath Introducer is intended to provide access and facilitate the percutaneous introduction of various devices into veins and/or arteries while maintaining hemostasis for a variety of diagnostic and therapeutic procedures. The device can also provide access to a native or synthetic graft used for hemodialysis. The side port of the sheath allows adequate flow to perform temporary hemodialysis. The device is not indicated for long term vascular or hemodialysis access.

Recall number
Z-1497-2017
Recalling firm
Merit Medical Systems, Inc.
Product
7F Prelude Short Sheath Introducer. Catalog Numbers: K15-00070, K15-00170, PSS-7F-4-035MT, PSS-7F-4MT. Produc Usage: The Merit Prelude Short Sheath Introducer is intended to provide access and facilitate the percutaneous introduction of various devices into veins and/or arteries while maintaining hemostasis for a variety of diagnostic and therapeutic procedures. The device can also provide access to a native or synthetic graft used for hemodialysis. The side port of the sheath allows adequate flow to perform temporary hemodialysis. The device is not indicated for long term vascular or hemodialysis access.
Status
Terminated
Initiated
2017-02-16
Posted
2017-03-24
Terminated
2017-08-08
Reason
Merit Medical Systems, Inc. announces a voluntary field action for sheaths included with the 7F Prelude Short Sheath Introducers because the sheath marker tips may detach during use.
Root cause
Process control
Action
Consignees notified on 2/16/17 to immediately stop using or distributing the affected lots and return them to Merit. For any questions, please contact your Merit Sales Representative or Merit Customer Service at 801.208.4381.
Quantity
1,265 total
Distribution
Nationwide and other foreign consignees. No Canadian or VA/govt/military consignees.
Lot, serial or model codes
Lot Numbers: H1041469, H1041473, H1036880, H1041464.
510(k) numbers
["K082063"]
PMA numbers
not on file
Product code
DYB
Firm FEI
1721504
Firm city
South Jordan
Firm state
UT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register