Wright Medical Technology, Inc.: CROSSCHECK(R) Plating system, Utility Plate, 5-Hole, REF CCP-UTN5. Orthopedic use.

Recall number
Z-1497-2015
Recalling firm
Wright Medical Technology, Inc.
Product
CROSSCHECK(R) Plating system, Utility Plate, 5-Hole, REF CCP-UTN5. Orthopedic use.
Status
Terminated
Initiated
2015-03-06
Posted
2015-04-22
Terminated
2018-08-05
Reason
The minor diameter, major diameter, thread pitch, and pitch on the Crosscheck are out of specification, which may result in intraoperative malfunction
Root cause
Nonconforming Material/Component
Action
Firm has initiated the recall on 03/06/2015 through written communication via Fed Ex P1 delivery to all Wright Medical Technologies distributors (hubs). Follow-up to non-responding consignees will commence 30 days after initiation. Effectiveness checks will be performed through the confirmation of returned product via a validated ERP system, a returned/signed verification form, and in the case of non-responders, a phone call, e-mail or follow up letter, depending on the region and availability of consignee.
Quantity
138 units
Distribution
Nationwide Distribution.
Lot, serial or model codes
Lot Numbers: 3504933, 35049-33A, 35049-33A-1, 35049-34, 362804, 36280-4-1-A
510(k) numbers
["K132594"]
PMA numbers
not on file
Product code
HRS
Firm FEI
3010667733
Firm city
Memphis
Firm state
TN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register