Skip to the content

FDA Medical Device Recalls (openFDA)

PF Consumer Healthcare 1 LLC: ThermaCare HEATWRAPS, ADVANCED BACK PAIN THERAPY, UPC 305733010037 - Product Usage: ThermaCare(R) HeatWraps are indicated in adults to provide heat therapy for the temporary relief of minor muscular and joint pains associated with overexertion, strains, sprains, and arthritis and menstrual pain. ThermaCare HeatWraps are made up of heat cells enclosed within wraps that consists of a mixture of ingredients including soft iron, carbon or absorbent gelling material and brine solution (water, salt, and sodium thiosulphate). The heat cells get activated in the presence of air (oxygen) to produce a low-level heat.

Recall number
Z-1495-2019
Recalling firm
PF Consumer Healthcare 1 LLC
Product
ThermaCare HEATWRAPS, ADVANCED BACK PAIN THERAPY, UPC 305733010037 - Product Usage: ThermaCare(R) HeatWraps are indicated in adults to provide heat therapy for the temporary relief of minor muscular and joint pains associated with overexertion, strains, sprains, and arthritis and menstrual pain. ThermaCare HeatWraps are made up of heat cells enclosed within wraps that consists of a mixture of ingredients including soft iron, carbon or absorbent gelling material and brine solution (water, salt, and sodium thiosulphate). The heat cells get activated in the presence of air (oxygen) to produce a low-level heat.
Status
Terminated
Initiated
2019-04-26
Posted
not on file
Terminated
2024-06-11
Reason
There is the potential that a HeatWrap could include cells that have a higher cell temperature than specified in lot S97473. The use of heat cell wrap with increased temperatures poses a potential risk of skin injuries such as burns/blisters and/or skin irritation on the wrap-applied area.
Root cause
Other
Action
The firm initiated the recall by press release on 04/26/2019. Notification packages to direct consignees were sent on 04/29/2019. The consignees were directed to cease distribution and return the product. Distributors were directed to sub-recall. Retailers were asked to post the press release either at the cash register or another prominent location.
Quantity
155,904 individual HeatWraps
Distribution
Worldwide distribution - US Nationwide and PR, and countries of Ireland, UK, Malta, Netherlands, Germany, and Switzerland.
Lot, serial or model codes
Lot Number S97473
510(k) numbers
not on file
PMA numbers
not on file
Product code
IMD
Firm FEI
3007593958
Firm city
Albany
Firm state
GA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register