Zimmer Biomet, Inc.: Orthosize Templating Version 1.2.6 Echo Bi-Metric Hip Stem Digital Templates. For preoperative planning of orthopedic surgery.

Recall number
Z-1495-2017
Recalling firm
Zimmer Biomet, Inc.
Product
Orthosize Templating Version 1.2.6 Echo Bi-Metric Hip Stem Digital Templates. For preoperative planning of orthopedic surgery.
Status
Terminated
Initiated
2017-02-14
Posted
2017-03-13
Terminated
2017-11-03
Reason
Digital templates were created with the incorrect files.
Root cause
Software Manufacturing/Software Deployment
Action
On 2/14/2017 URGENT MEDICAL DEVICE CORRECTION notifications were sent to the affected users and 3rd party users via email. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for 3rd party users to respond to the formal recall notification. Orthosize Templating Users In approximately 6 weeks you will receive a notice that an update is available for the Orthosize Templating App. Upon selecting the app you will be required to update the app in order to use it. 1. Review this notification and ensure affected personnel are aware its contents. 2. Immediately update the app when the update becomes available. 3. If after reviewing this notice you have further questions or concerns please call 411 Technical Services at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Alternatively, your questions may be sent by email to corporatequality.postmarket@zimmerbiomet.com Digital Template Users (3rd party) 1. Review this notification and ensure affected personnel are aware of the contents. 2. Immediately update your systems to include the revised digital templates included with notice. 3. Complete Attachment 1 Certificate of Acknowledgement. a. Return a digital copy to corporatequality.postmarket@zimmerbiomet.com. b. Retain a copy of the Acknowledgement Form with your records. 4. If after reviewing this notice you have further questions or concerns please call 411 Technical Services at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of the call center operating hours will receive a prompt to record a voicemail or be transferred to an on-call representative in the case of an emergency. Alternatively, your questions may be sent by email to corporatequality.postmarket@zimmerbiomet.com
Quantity
3232
Distribution
Worldwide distribution. US Nationwide including Puerto Rico, Canada, Argentina, Australia, Austria, Belgium, Chile, China, Denmark, Finland, France, Germany, Greece, India, Italy, Japan, Korea, Lithuania, Malaysia, Mexico, Netherlands, Nicaragua, Norway, Pakistan, Poland, Portugal, Romania, Singapore, South Africa, Spain, Sweden, Switzerland, Turkey, United Kingdom.
Lot, serial or model codes
Orthosize Templating Version 1.2.6
510(k) numbers
["K120115"]
PMA numbers
not on file
Product code
LLZ
Firm FEI
1825034
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register