Bio-Rad Laboratories Inc: PhD System EIA/IFAversion 2.1A software, model number 426-0246, a component of the Bio-Rad PhD System. Manufactured by Bio-Rad Laboratories, France, Distributed by Bio-Rad Laboratories, Hercules, CA. EIA and IFA processing system that consists of a PhD fluid handling station (s) linked to a network computer via ethernet hub. The computer provides worklist generation, data management, data reduction and microplate reader control functions.

Recall number
Z-1495-2011
Recalling firm
Bio-Rad Laboratories Inc
Product
PhD System EIA/IFAversion 2.1A software, model number 426-0246, a component of the Bio-Rad PhD System. Manufactured by Bio-Rad Laboratories, France, Distributed by Bio-Rad Laboratories, Hercules, CA. EIA and IFA processing system that consists of a PhD fluid handling station (s) linked to a network computer via ethernet hub. The computer provides worklist generation, data management, data reduction and microplate reader control functions.
Status
Terminated
Initiated
2009-09-29
Posted
2011-03-02
Terminated
2011-03-11
Reason
Incorrect patient results are being generated from out-of-range samples tested withc ompetitive enzyme immunoassays on the PhD system with software version 2.1A.
Root cause
Software change control
Action
BIO-RAD sent an URGENT: MEDICAL DEVICE CORRECTION letter dated September 29, 2009, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to immediately examine their customer records to identify all PhD System customers that are using version 2.1A of the EIA/IFA Software and Methods CD-ROM. Customers who are not running competitive immunoassays on the PhD System may continue using the 2.1A software. Customers who run competitive immunoassays must revert to software version 2.0B. Customers were instructed to translate if necessary and add local contact information to the customer letter and distribute to all PhD System customers that are using 2.1A of the EIA/IFA Software and Methods CD-ROM. A Customer Medical Device Correction Response Form must be collected from each customer to ensure that they have received this important communication. Once forms have been collected from all applicable customers, they should be faxed to Bio-Rad CSD Regulatory Affairs Department along with the completed Subsidiary Medical Device Correction Response Form. For questions regarding this recall call 510-741-3954.
Quantity
262 units
Distribution
Worldwide Distribution - USA including AL, CA, CO, CT, DC, FL, IL, MA, MD, MN, MO, MS, MT, NC, NY, OH, PA, SD, TN, TX, UT, and VT and the countries of Australia, New Zealand, Singapore, China, France, Sweden, and Canada
Lot, serial or model codes
Model 426-0246, software version 2.1A
510(k) numbers
not on file
PMA numbers
not on file
Product code
JQW
Firm FEI
2915274
Firm city
Hercules
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register