FDA Medical Device Recalls (openFDA)

Olympus Corporation of the Americas: Brand Name: Olympus ShockPulse-SE Lithotripsy System with Generator Product Name: ShockPulse-SE Lithotripsy System - Reuseable Probes Model/Catalog Number: SPL-SR, contains generator SPL-G Product Description: An assembly of devices that uses a combination of, or individually applied, pneumatically-created ballistic shock waves and ultrasonic waves for the intracorporeal fragmentation of stones (calculi) in the urinary tract (i.e., kidney, ureter, and bladder) for their removal. It typically includes a pressure and power regulator/control unit, handpieces, and two types of probes of various sizes. The probes are attached to the two handpieces which are connected to the control unit that supplies the energy for both the pneumatic and ultrasonic functions. Energy is applied to a calculus which is fragmented in situ and removed via a suction system facilitated by the lithotripsy system.

Recall number
Z-1494-2026
Recalling firm
Olympus Corporation of the Americas
Product
Brand Name: Olympus ShockPulse-SE Lithotripsy System with Generator Product Name: ShockPulse-SE Lithotripsy System - Reuseable Probes Model/Catalog Number: SPL-SR, contains generator SPL-G Product Description: An assembly of devices that uses a combination of, or individually applied, pneumatically-created ballistic shock waves and ultrasonic waves for the intracorporeal fragmentation of stones (calculi) in the urinary tract (i.e., kidney, ureter, and bladder) for their removal. It typically includes a pressure and power regulator/control unit, handpieces, and two types of probes of various sizes. The probes are attached to the two handpieces which are connected to the control unit that supplies the energy for both the pneumatic and ultrasonic functions. Energy is applied to a calculus which is fragmented in situ and removed via a suction system facilitated by the lithotripsy system.
Status
Open, Classified
Initiated
2026-01-08
Posted
2026-03-05
Terminated
not on file
Reason
Mis-wired component-the improperly wired component can result in additional noise on the power supply output. Noise on the ultrasonic input circuit is not consistent with electromagnetic compatibility (EMC) standards and may decrease the overall reliability of the system and result in delay of treatment. It was discovered that additional serial numbers of ShockPulse generators may also have the mis-wired component. As a result, this is expanding the action to the additional impacted devices.
Root cause
Device Design
Action
On January 8, 2026, URGENT: MEDICAL DEVICE CORRECTION letters were sent to customers. Actions to be taken: 1. Examine your inventory and identify any affected devices. 2. Contact your Olympus Customer Service contact to obtain a Return Material Authorization and arrange for the return of your device. Olympus will repair the affected part at no charge and return the device back to you. 3. If you have further distributed the affected products, please forward this notification to other users who may have them. Olympus requests you to report any complaints related to the ShockPulse device to our Technical Assistance Center (TAC) at 1- 800-848-9024, option 1, and the FDA. Adverse events experienced with the use of this product may also be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax.
Quantity
55 units (All OUS)
Distribution
International distribution in the country of Canada, Germany, Singapore, Australia, and India.
Lot, serial or model codes
Model/Catalog Number: SPL-SR, contains generator SPL-G; UDI: SPL-SR 00821925043824, SPL-G 00821925044203; Serial Numbers: CG5026, CG5027, CG5028, CG5029, CG5030, CG5031, CG5032, CG5033, CG5034, CG5035, CG5036, CG5037, CG5039, CG5040, CG5041, CG5042, CG5043, CG5045, CG5046, CG5047, CG5048, CG5049, CG5050, CG5051, CG5052, CG5053, CG5054, CG5055, CG5056, CG5057, CG5058, CG5059, CG5060, CG5061, CG5062, CG5063, CG5065, CG5066, CG5067, CG6037, CG6061, CG6069, CG6070, CG6071, CG6088, CG6094, CG6095, CG6096, CG6097, CG6098, CG6099, CG7000, CG7001, CG7002, CG7003;
510(k) numbers
["K171024"]
PMA numbers
not on file
Product code
FEO
Firm FEI
2429304
Firm city
Center Valley
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register