Skip to the content

FDA Medical Device Recalls (openFDA)

Zimmer Surgical Inc: Zimmer Biomet Alvarado Knee Holder Base Plate Assembly and Foot Piece, & Alvarado II Base Plate and Foot Piece: Alvarado Foot Piece: Item Number: 00-1320-010-00 Alvarado Knee Holder Base Plate Assembly: Item Number: 00-1320-011-00 Alvarado II Foot Piece: Item Number: 00-1320-210-00 Alvarado II Base Plate: Item Number: 00-1320-211-00 Product Usage: This device is designed to hold the patient s knee in the selected position during surgery. The Alvarado Knee Holder includes a Base Plate assembly with two rows of slotted hooks. These rows extend upward from the base plate for a sufficient length to provide adequate selective adjustment. A Foot Piece is mounted on an alignment rod that is pivotally engaged into the hook portion of the slots.

Recall number
Z-1494-2019
Recalling firm
Zimmer Surgical Inc
Product
Zimmer Biomet Alvarado Knee Holder Base Plate Assembly and Foot Piece, & Alvarado II Base Plate and Foot Piece: Alvarado Foot Piece: Item Number: 00-1320-010-00 Alvarado Knee Holder Base Plate Assembly: Item Number: 00-1320-011-00 Alvarado II Foot Piece: Item Number: 00-1320-210-00 Alvarado II Base Plate: Item Number: 00-1320-211-00 Product Usage: This device is designed to hold the patient s knee in the selected position during surgery. The Alvarado Knee Holder includes a Base Plate assembly with two rows of slotted hooks. These rows extend upward from the base plate for a sufficient length to provide adequate selective adjustment. A Foot Piece is mounted on an alignment rod that is pivotally engaged into the hook portion of the slots.
Status
Terminated
Initiated
2019-04-16
Posted
not on file
Terminated
2020-05-19
Reason
Foreign Material in the Alvarado Knee Holder Base Plate Assembly and Foot Piece and Alvarado II Base Plate and Foot Piece due to potentially inadequate cleaning procedures
Root cause
Device Design
Action
Zimmer Biomet issued letter on 4/10/19 via email to distributors; and Hospital risk managers, as well as distributors with product notified via courier. Distributors letter identifying the issue and their responsibilities. These responsibilities include locating and removing the product in their territory, as well as identifying hospitals who have unconsumed inventory on hand. Distributors will return on-hand product to Zimmer Biomet and ensure all of their products are accounted for using the form provided in the letter. Hospital risk managers provided with a letter identifying the issue and their responsibilities. These responsibilities include: - Locating and returning the product in their territory. - Returning Attachment 1 (Certificate of Acknowledgement) to Zimmer Biomet. Questions or concerns after reviewing this notice, please call customer service at 800-830-0970 between 8:00 am and 5:00pm EST, Monday through Friday. Updates to the FAQ (clarification) sent via email on April 19 to distributors.
Quantity
37,402
Distribution
Worldwide distribution - US Nationwide and countries of AUKLAND AUSTRALIA BRAZIL CANADA CHILE Dominican Republic EL SALVADOR GUATEMALA HONG KONG INDIA JAPAN MEXICO NETHERLANDS NEW ZEALAND KOREA SINGAPORE TAIWAN VENEZUELA
Lot, serial or model codes
All lot codes
510(k) numbers
not on file
PMA numbers
not on file
Product code
JEB
Firm FEI
1526350
Firm city
Dover
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register