Skip to the content

FDA Medical Device Recalls (openFDA)

Nidek Inc: Final Fit Software Version 1.11 and 1.12; PC Based software installed outside Nidek EC-5000 Excimer Laser System. Ophthalmic laser system use for correction of corneal refraction and ablation of the corneal surface. The system is composed of a laser generator which produces an excimer laser radiation of wavelength 193nm, a beam delivery unit, an optical system for observation, a gas system and a computer for system control.

Recall number
Z-1494-2017
Recalling firm
Nidek Inc
Product
Final Fit Software Version 1.11 and 1.12; PC Based software installed outside Nidek EC-5000 Excimer Laser System. Ophthalmic laser system use for correction of corneal refraction and ablation of the corneal surface. The system is composed of a laser generator which produces an excimer laser radiation of wavelength 193nm, a beam delivery unit, an optical system for observation, a gas system and a computer for system control.
Status
Terminated
Initiated
2017-02-01
Posted
not on file
Terminated
2017-12-05
Reason
During treatment planning, the procedure was programmed with an unintended (wrong) correction.
Root cause
Software change control
Action
Nidek sent a Recall notification letter dated February 1, 2017 to all affected consignees by certified mail method beginning February 2, 2017. Letters advised of the issue received from a complaint and the steps the user took that created the issue. To prevent the issue, the letters provide steps that should be taken. Nidek's Customer Support team will contact customers by phone or email. Customers with questions were instructed to contact Customer Service at 1-800-223-9044 or email ff_support@nidek.com.
Quantity
29
Distribution
Nationwide Distribution to MN, NV, OH, NC, CA, CO, CA, GA, VA, MI, AZ, PA, TN, WA, TX, NY.
Lot, serial or model codes
Software version 1.11 and 1.12. Serial numbers of Affected devices: MNE0009; NVE0002; OH00295; NC00185; CAE0016; COE0008; MI00131; GA00241; CAE0007; VAE0004; MI00131; CAE0031; AZ00167; COE0011; CA00561; PAE0003; TN00168; MNE0009; WA00246; CAE0047; OHE0005; WAE0008; MI00244; NY00473; CAE0007; WAE0008; NVE0002; WA00233; CAE0016.
510(k) numbers
not on file
PMA numbers
["P970053"]
Product code
LZS
Firm FEI
2936921
Firm city
Fremont
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register