Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc: Siemens-branded ONCOR" or PRIMUS" linear accelerator with an OPTIVUE" or MVision" imaging system; Model Numbers: ONCOR Expression 7360717, ONCOR Impression Plus 5857912, ONCOR Avant Garde 5863472, PRIMUS HI 4504200; ARTISTE/ONCOR/PRIMUS systems are manufactured by: SIEMENS AG, MEDICAL SOLUTIONS, Roentgenstrasse 19-21, Kemnath, GERMANY 95478 Product Usage: Medical charged-particle radiation therapy system. The intended use of the SIEMENS branded ONCOR and PRIMUS family of linear accelerator systems is to deliver X-ray photon and electron radiation for the therapeutic treatment of cancer.

Recall number
Z-1494-2012
Recalling firm
Siemens Medical Solutions USA, Inc
Product
Siemens-branded ONCOR" or PRIMUS" linear accelerator with an OPTIVUE" or MVision" imaging system; Model Numbers: ONCOR Expression 7360717, ONCOR Impression Plus 5857912, ONCOR Avant Garde 5863472, PRIMUS HI 4504200; ARTISTE/ONCOR/PRIMUS systems are manufactured by: SIEMENS AG, MEDICAL SOLUTIONS, Roentgenstrasse 19-21, Kemnath, GERMANY 95478 Product Usage: Medical charged-particle radiation therapy system. The intended use of the SIEMENS branded ONCOR and PRIMUS family of linear accelerator systems is to deliver X-ray photon and electron radiation for the therapeutic treatment of cancer.
Status
Terminated
Initiated
2012-04-04
Posted
2012-05-08
Terminated
2012-09-17
Reason
The firm became aware of a potential issue with Siemens-branded ONCOR or PRIMUS linear accelerator with an OPTIVUE or MVision imaging system that has the potential collision problem of the imaging panel if the panel does not retract before the gantry rotates; this may result in patient injury.
Root cause
Process control
Action
Siemens sent an Update Instruction containing the Customer Information letters beginning April 4, 2012 delivered by Siemens customer Service Group or by certified mail to all affected customers. The letter identified the affected product, what is the issue and what has Siemens done to address this issue. The letter instructs customer to include this Customer Information in their Digital Linear Accelerator System Owner Manual chapter Safety Advisory Letters where it should remain.
Quantity
73 units
Distribution
Worldwide Distribution - USA Nationwide
Lot, serial or model codes
All units of these models.
510(k) numbers
["K060226","K993425"]
PMA numbers
not on file
Product code
IYE
Firm FEI
2937457
Firm city
Concord
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register