FDA Medical Device Recalls (openFDA)
Siemens Medical Solutions USA, Inc: ARTISTE or ONCOR Linear Accelerators in combination with 160MLC (Multi Leaf Collimator). Linear Accelerators that may be affected are: ARTISTE MV System; ONCOR Avant Garde ONCOR Impression ONCOR Impression Plus ONCOR Expression. Manufactured by Siemens AG, Medical Solutions, Kemnath, Germany; Registered by Siemens Medical Solutions, Concord, CA. ARTISTE and ONCOR family of Linear Accelerators are used to deliver X-ray photon and electron radiation for the therapeutic treatment of cancer. Product Model with Part Number: ARTISTE MV System, 8139789; ONCOR Avant Garde, 5863472; ONCOR Impression, 5857920; ONCOR Impression, Plus 5857912; ONCOR Expression, 7360204; ONCOR Expression, 7360717. ARTISTE and ONCOR family of Linear Accelerators are used to deliver X-ray photon and electron radiation for the therapeutic treatment of cancer.
- Recall number
- Z-1493-2012
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Product
- ARTISTE or ONCOR Linear Accelerators in combination with 160MLC (Multi Leaf Collimator). Linear Accelerators that may be affected are: ARTISTE MV System; ONCOR Avant Garde ONCOR Impression ONCOR Impression Plus ONCOR Expression. Manufactured by Siemens AG, Medical Solutions, Kemnath, Germany; Registered by Siemens Medical Solutions, Concord, CA. ARTISTE and ONCOR family of Linear Accelerators are used to deliver X-ray photon and electron radiation for the therapeutic treatment of cancer. Product Model with Part Number: ARTISTE MV System, 8139789; ONCOR Avant Garde, 5863472; ONCOR Impression, 5857920; ONCOR Impression, Plus 5857912; ONCOR Expression, 7360204; ONCOR Expression, 7360717. ARTISTE and ONCOR family of Linear Accelerators are used to deliver X-ray photon and electron radiation for the therapeutic treatment of cancer.
- Status
- Terminated
- Initiated
- 2012-04-03
- Posted
- 2012-05-07
- Terminated
- 2012-07-25
- Reason
- Siemens became aware of a potential safety risk related to the behavior of the 160 MLC when using the "Motion Stop" button followed by "Reset" several times during the treatment of one beam or one segment (issue 1). Additionally, there may be a potential for collision if the operator does not ensure clearance of the extended flat panel positioner for all gantry rotation angles (issue 2).
- Root cause
- Labeling design
- Action
- Siemens sent a Safety Advisory Notice with Update to all affected customers. The notice informed the customers of the risk of pressing the Motion Stop button repeatedly during treatment. Work Steps were provided for customers to follow.
- Quantity
- not on file
- Distribution
- Worldwide Distribution - USA including the states of CA, DE, FL, IL, KY, LA, MA, MI, NE, NJ, NY, OH, PA, SD, TN, UT, WI, and WV
- Lot, serial or model codes
- Product Model with Part Number: ARTISTE MV System, 8139789; ONCOR Avant Garde, 5863472; ONCOR Impression, 5857920; ONCOR Impression, Plus 5857912; ONCOR Expression, 7360204; ONCOR Expression, 7360717
- 510(k) numbers
- ["K072486","K092145"]
- PMA numbers
- not on file
- Product code
- IYE
- Firm FEI
- 2937457
- Firm city
- Concord
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as