Skip to the content

FDA Medical Device Recalls (openFDA)

ConMed Corporation: Neotrode Wire Attached, Catalog Number: 1731-003; Neotrode II Wire Attached, Catalog Number: 1741-003; Softrace Neonatal Wire Attached, Catalog Number: 2321-003; Softrace Pediatric Wire Attached, Catalog Number: 2331-003; Softrace Limb Band, Catalog Number: 2310-003; TRU-LINK ECG Electrode (Neonatal Pre-Wired), Catalog Number: 685-0037-00; for use with electrocardiographic monitoring equipment to detect neonatal/pediatric heart action voltages.

Recall number
Z-1493-2011
Recalling firm
ConMed Corporation
Product
Neotrode Wire Attached, Catalog Number: 1731-003; Neotrode II Wire Attached, Catalog Number: 1741-003; Softrace Neonatal Wire Attached, Catalog Number: 2321-003; Softrace Pediatric Wire Attached, Catalog Number: 2331-003; Softrace Limb Band, Catalog Number: 2310-003; TRU-LINK ECG Electrode (Neonatal Pre-Wired), Catalog Number: 685-0037-00; for use with electrocardiographic monitoring equipment to detect neonatal/pediatric heart action voltages.
Status
Terminated
Initiated
2010-09-23
Posted
2011-03-02
Terminated
2014-10-28
Reason
Poor signal or loss of signal. Certain recalled devices have a loose wire attachment that may lead to intermittent or poor signal quality.
Root cause
Nonconforming Material/Component
Action
ConMed Corporation distributed an Urgent Device Recall notification letter on Tru-Link ECG Electrode (Neonatal Pre-wired) dated September 23, 2010. ConMed Corp. has advised to cease the use of the devices immediately and conducted a recall to the user level. Direct consignees were asked to contact sub-accounts if recalled devices were further distributed and notified them of such recall. Recalled devices were to return to ConMed Corp. at 525 French Road, Utica, NY 13502 with completed Attachment II to indicate the amount returned. Customers with no recalled devices were asked to fill out Attachment II and faxed to 315-624-3225. Questions can be directed to Ms. Patricia Cotter, ConMed Recall Coordinator at 315-624-3237 or fax to 315-624-3225 or email to neorecall@conmed.com.
Quantity
9,009,000 Pediatric/Neonatal ECG Electrodes with Attached Wire
Distribution
Worldwide distribution, including USA, Dubai, Ecuador, Egypt, Germany, Holland, Indonesia, Israel, Italy, Jamaica, Kenya, Korea, Lithuania, Malaysia, Mexico, New Zealand, Norway, Qatar, Saudi Arabia, Singapore, South Africa, Sweden, Switzerland, US Virgin Islands, Canada and Belgium.
Lot, serial or model codes
Neotrode, Wire Attached (Catalog Number: 1731-003) Lot Code: 0904014 thru 1008134; Neotrode II, Wire Attached (Catalog Number: 1741-003) Lot Code: 0904014 thru 1008114; Softrace Limb Band (Catalog Number: 2310-003) Lot Code: 0904014 thru 1008274; Softrace Neonatal, Wire Attached (Catalog Number: 2321-003) Lot Code: 0904014 thru 1007234; Softrace Pediatric, Wire Attached (Catalog Number: 2331-003) Lot Code: 0904014 thru 1007214; and TRU-LINK ECG Electrodes (Neonatal Pre-Wired) (Catalog Number: 685-0037-00) Lot Code: 0904014 thru 1007064
510(k) numbers
["K901555"]
PMA numbers
not on file
Product code
DRX
Firm FEI
3001451571
Firm city
Utica
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register