FDA Medical Device Recalls (openFDA)
Diagnostica Stago, Inc.: Brand Name: STA Liatest Free Protein S Product Name: STA Liatest Free Protein S Model/Catalog Number (REF): 00516 Product Description: The STA - Liatest Free Protein S kits are intended for use on STA-R and STA Compact for the quantitative antigenic assay of free protein S in human citrated plasma by the immuno-turbidimetric method. Component: Not Applicable
- Recall number
- Z-1491-2026
- Recalling firm
- Diagnostica Stago, Inc.
- Product
- Brand Name: STA Liatest Free Protein S Product Name: STA Liatest Free Protein S Model/Catalog Number (REF): 00516 Product Description: The STA - Liatest Free Protein S kits are intended for use on STA-R and STA Compact for the quantitative antigenic assay of free protein S in human citrated plasma by the immuno-turbidimetric method. Component: Not Applicable
- Status
- Open, Classified
- Initiated
- 2026-02-04
- Posted
- 2026-03-03
- Terminated
- not on file
- Reason
- The potential of out-of-range results and an underestimation of the free protein S level in normal patient plasmas. The defect only affects normal-range values.
- Root cause
- Under Investigation by firm
- Action
- On February 4, 2026 URGENT: MEDICAL DEVICE RECALL (REMOVAL) letters were sent to customers. Actions required: If you currently have one or more kits of STA LIATEST FREE PROTEIN S 6 : - For REF 00516: Lots 271971, 272859, and 273018 with a short expiration date we ask you to discontinue their use and dispose of them in accordance with your local regulations. Stago will provide a credit for any destroyed kits. - For REF 00516: lot 273426: o If the initial barcode sheet provided with the kit has already been loaded into the instrument: - Stop the use the initial barcode pre-calibration settings. - Replace the initial barcode settings with the new pre-calibration barcode settings attached, by following to the procedure described in Appendix 1 of this letter. o If the initial barcode sheet has not yet been loaded into the instrument: - Do not use it and proceed to load the lot using the new pre-calibration barcodes provided in the attachment.
- Quantity
- 369
- Distribution
- Worldwide - US Nationwide distribution in the states of AZ, CA, CO, FL, GA, HI, IL, KY, MI, MN, MO, MT, NC, ND, NJ, NY, OH, PA, TN, TX, UT, VA, WA and the countries of ALBANIA, ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BAHRAIN, BANGLADESH, BELGIUM, BIELORUSSIE, BRAZIL, BULGARIA, CANADA, CHILE, CHINA, COSTA RICA, DENMARK, FINLAND, FRANCE, GERMANY, GREECE, GUADELOUPE, GUATEMALA, INDIA, IRAQ, IRELAND, ITALY, JORDAN, KENYA, KOREA REPUBLIC OF, LITUANIA, LUXEMBOURG, MALAYSIA, MEXICO, MOROCCO, NETHERLANDS, NEW CALEDONIA, NEW ZEALAND, PAKISTAN, PALESTINIAN TERRITORY, PARAGUAY, PORTUGAL, ROMANIA, RUSSIE, SAUDI ARABIA, SERBIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TUNISIE, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, URUGUAY;
- Lot, serial or model codes
- Model/Catalog Number (REF): 00516; UDI: (1) (01)03607450005165(11)241001(17)260131(10)271971(241)00516, (2) (01)03607450005165(11)250221(17)260531(10)272859(241)00516, (3) (01)03607450005165(11)250306(17)260630(10)273018(241)00516, (4) (01)03607450005165(11)250430(17)260831(10)273426(241)00516; Lot numbers: (1) 271 971, (2) 272 859, (3) 273 018, (4) 273 426; Lot Code: 271 971 (01)03607450005165(11)241001(17)260131(10)271971(241)00516
- 510(k) numbers
- ["K010963"]
- PMA numbers
- not on file
- Product code
- GGP
- Firm FEI
- 2245451
- Firm city
- Parsippany
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as