Skip to the content

FDA Medical Device Recalls (openFDA)

Olympus Corporation of the Americas: Olympus Single Use Biopsy Valve. Model/Catalog Number: MAJ-1218. 20 units per box. The MAJ-1218 has been designed to be attached to the instrument channel port of compatible endoscopes to prevent leakage of body fluids.

Recall number
Z-1489-2026
Recalling firm
Olympus Corporation of the Americas
Product
Olympus Single Use Biopsy Valve. Model/Catalog Number: MAJ-1218. 20 units per box. The MAJ-1218 has been designed to be attached to the instrument channel port of compatible endoscopes to prevent leakage of body fluids.
Status
Open, Classified
Initiated
2026-01-29
Posted
2026-03-03
Terminated
not on file
Reason
Potential for rubber fragment detachment during use.
Root cause
Under Investigation by firm
Action
Olympus notified consignees on about 01/29/2026 via letter titled URGENT: FIELD SAFETY NOTICE. Consignees were instructed to carefully read the content of the notification, ensure all personnel are completely knowledgeable and thoroughly trained on the content of this notification, and that they may continue using the devices as per the notification and instructions for use. Olympus is highlighting the importance of strictly adhering to the Inspection of the Biopsy Valve and Inserting and Withdrawing the Endo-therapy accessories sections in the IFU related to the detachment of fragment from the slit part of rubber. Consignees were requested to acknowledge receipt of the notification via their portal and if product has been further distributed to notify those consignees.
Quantity
1,276 boxes (20 units per box)
Distribution
US Nationwide distribution.
Lot, serial or model codes
Model/Catalog Number: MAJ-1218. UDI-DI: 14953170154291. All Lot Numbers. 20 units per box.
510(k) numbers
["K061313"]
PMA numbers
not on file
Product code
EOQ
Firm FEI
2429304
Firm city
Center Valley
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register