Skip to the content

FDA Medical Device Recalls (openFDA)

GE Healthcare: GE Healthcare Centricity Centricity PACS-IW with Universal Viewer, Model Numbers: 1) 2049587-015; 2) 2066908-086; 3) 2066908-136; 4) 2104867-045.

Recall number
Z-1487-2025
Recalling firm
GE Healthcare
Product
GE Healthcare Centricity Centricity PACS-IW with Universal Viewer, Model Numbers: 1) 2049587-015; 2) 2066908-086; 3) 2066908-136; 4) 2104867-045.
Status
Open, Classified
Initiated
2025-02-19
Posted
2025-03-28
Terminated
not on file
Reason
There is a potential security vulnerability in Centricity Universal Viewer (UV), Centricity PACS-IW (PACS-IW), Centricity Radiology RA600 (RA600), & Centricity Cardiology CA1000 (CA1000) where the service login credentials are able to be identified which could allow a malicious actor with these credentials to access the system and potentially manipulate patient data.
Root cause
Software design
Action
GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 02/19/2025 by letter delivered using traceable means. The notice explained the issue, safety issue, and requested the following actions be taken: "You can continue to use your device. Please ensure all potential users in your facility are made aware of this safety notification. Please ensure you have implemented the actions below: 1: Ensure your hospital network is secure and 2: Outside of the hospital network implement additional safeguards such as a VPN. Please retain this document for your records. Please complete and return the attached acknowledgement form to recall.85480@gehealthcare.com." "If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative." Per status report 05/23/2025, the firm send a follow-up notice dated 05/13/2025 providing an installation manual addendum that provides instructions for changing passwords.
Quantity
4 units
Distribution
Worldwide distribution.
Lot, serial or model codes
Model/catalog/code # 2049587-015, Installed Product ID 100707-1-Centricity Universal Viewer-00558649, UDI UDI Not Required, System ID PACS-IW-126; Model/catalog/code # 2066908-086, Installed Product ID 9842-4-Centricity Universal Viewer-19386293, UDI UDI Not Required, System ID GA001PACTRL01; Model/catalog/code # 2066908-136, Installed Product ID ZA2533-Centricity Universal Viewer-00116577, UDI UDI Not Required, System ID ZA2533PAC1; Model/catalog/code # 2104867-045, Installed Product ID X9003062 - HC3062-Centricity Universal Viewer-01462397, UDI (01)00840682103800(10)6.0SP50412094097001D, System ID HC3062UV01.
510(k) numbers
not on file
PMA numbers
not on file
Product code
LLZ
Firm FEI
3004526608
Firm city
Chicago
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register