Philips North America, LLC: Integris Allura 15 & 12, monoplane (system code 722043) and biplane (system code 722044)
- Recall number
- Z-1487-2020
- Recalling firm
- Philips North America, LLC
- Product
- Integris Allura 15 & 12, monoplane (system code 722043) and biplane (system code 722044)
- Status
- Terminated
- Initiated
- 2019-12-30
- Posted
- not on file
- Terminated
- 2022-11-17
- Reason
- A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible.
- Root cause
- Component design/selection
- Action
- On December 30, 2019, the firm notified customers via Medical Device Correction letter. Customers were informed of the product issue and advised to do the following: If the capacitor of the Velara X-Ray fails as described (capacitor failure after large number of surges in a short period of time), please switch off the system, take it out of service immediately, and contact your local Philips representative. The firm will be resolving the issue by replacing a Printed Circuit Board (PCB) in the convertor, which prevents the capacitor from failing. PCB replacements started in January 2020. Customers will be contacted by their local Philips representatives to schedule a date for the PCB replacement. Customers with further questions should contact the Technical Support Line: 1-800-722-9377.
- Quantity
- 70
- Distribution
- Nationwide domestic distribution. Foreign distribution worldwide.
- Lot, serial or model codes
- Serial Codes 3217017 200 58 72 CV.170 180 524615 47 211 364 30 3862570/0011 529 S.N. 519 S01H060871 505/0050114 S01H057274 01H077213 530 532 S01HA52566 348 476 320 481 467 284 352 359 459 73*004 733 536 CV.474 454 49 281 51 186 122 250 37 22 CV.312 296 373 CV.596 176 30 580 279
- 510(k) numbers
- ["K002016"]
- PMA numbers
- not on file
- Product code
- OWB
- Firm FEI
- 1218950
- Firm city
- Andover
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as