Skip to the content

FDA Medical Device Recalls (openFDA)

Integra LifeSciences Corp.: TCC-EZ, Total Contact Cast System, Model No. TCC2-4051

Recall number
Z-1487-2018
Recalling firm
Integra LifeSciences Corp.
Product
TCC-EZ, Total Contact Cast System, Model No. TCC2-4051
Status
Terminated
Initiated
2017-08-18
Posted
not on file
Terminated
2018-07-19
Reason
A quarantined lot of the devices was released for sale. Devices from this lot may experience increased exothermic reaction causing higher levels of heat output which exceed the product specifications.
Root cause
Process control
Action
On 8/18/17, the recalling firm's sales reps notified affected consignees of the recall and distributed customer letters. Customers were asked to examine their inventory for product from the affected lot, warned not to use the product, complete the included "Acknowledgement and Return Form", and return the form by email or fax. Customers were provided with a Customer Service number: 1-800-654-2873.
Quantity
250
Distribution
US nationwide distribution.
Lot, serial or model codes
Lot 1216033
510(k) numbers
not on file
PMA numbers
not on file
Product code
ITG
Firm FEI
3003418325
Firm city
Plainsboro
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register