FDA Medical Device Recalls (openFDA)
AGFA Corp.: AGFA Digital Radiography X-Ray System DX-D 100 Agfa's DX-D 100 is indicated for use in providing diagnostic quality images to aid the physician with diagnosis. Systems can be used with MUSICA2 image processing to create radiographic images of the skeleton including skull, spinal column and extremities: chest, abdomen, and other body parts.
- Recall number
- Z-1487-2013
- Recalling firm
- AGFA Corp.
- Product
- AGFA Digital Radiography X-Ray System DX-D 100 Agfa's DX-D 100 is indicated for use in providing diagnostic quality images to aid the physician with diagnosis. Systems can be used with MUSICA2 image processing to create radiographic images of the skeleton including skull, spinal column and extremities: chest, abdomen, and other body parts.
- Status
- Terminated
- Initiated
- 2013-05-15
- Posted
- 2013-06-06
- Terminated
- 2014-12-31
- Reason
- Customers could potentially experience intermittent, unintended and illogical movement when using the product.
- Root cause
- Under Investigation by firm
- Action
- Agfa sent an "URGENT SAFETY NOTICE" letter dated May 15, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customers. Customers were instructed if you experience unintended movement, the unit has to be taken out of operation. Please notify you Agfa service contact at once. Please distribute this information within your facility to all those who need to be aware of this. Please complete the feedback form as soon as possible and return to us. For further questions please call 1-877-777-2432.
- Quantity
- 104
- Distribution
- Worldwide Distribution - US Nationwide including the states of AL, FL, IL, IN, IA, KY, MA, MI, MN, MS, MO, NE, NJ, NY, NC, OH, OR, SC, TN, VA, WV, WY and the country of Canada
- Lot, serial or model codes
- Material numbers in US - 5Q82F, 5R75Q, 5Q8ZB, 5R77U & 58Q1D - Material numbers in Canada - 5R78W & 5R77U
- 510(k) numbers
- ["K103597"]
- PMA numbers
- not on file
- Product code
- IZL
- Firm FEI
- 3001236302
- Firm city
- Greenville
- Firm state
- SC
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28