Skip to the content

FDA Medical Device Recalls (openFDA)

Philips Medical Systems: IntelliVue MultiMeasurement Server (MMS)/Multimeasurement Server, physiological patient monitoring system. Model: M3001A

Recall number
Z-1487-06
Recalling firm
Philips Medical Systems
Product
IntelliVue MultiMeasurement Server (MMS)/Multimeasurement Server, physiological patient monitoring system. Model: M3001A
Status
Terminated
Initiated
2006-07-24
Posted
2006-09-12
Terminated
2020-08-11
Reason
Patient monitor may display inaccurate reading when the Disposable Sp02 Sensor is not attached
Root cause
Other
Action
Philips is using the contract services of Sterling, Indianapolis, IN to contact all consignees regarding the correction/upgrade to the device. The letter notification dated 7/21/06 was mailed UPS Delivery Tracking the week of 07/24/06. Additionally, Philips included with the Urgent Device Correction Notice a procedure to mitigate risk until the upgrade is installed.
Quantity
48,699 units
Distribution
Worldwide, including USA, Canada, Argentina, Australia, Austria, Belgium, Brazil, Czech, Croatia, China, Denmark, Finland, France, Germany, Israel, Italy, Jordan, Kuwait, Luxemburg, Netherlands, Oman, Norway, Poland, Portugual, Poland, Russia, South Africa, Spain, Sweden, Swizerland, Turkey, UAE, Yemen, UK, Zimbabwe, Zambia, India, Indonesia, Japan, Korea, Philipines, Malaysia, Singapore, Sri Lanka, Taiwan, Vietnam, Venezuela.
Lot, serial or model codes
M3001A options A01, A01C06, A01C12, and A01C18 with serial number prefix DE512 and MMS software revision C.0 or B.1 M3001A options A02, A02C06, and A02C18 with serial number prefix DE441 and DE512 and MMS software revision C.0 or B.1
510(k) numbers
["K033715"]
PMA numbers
not on file
Product code
DQA
Firm FEI
1218950
Firm city
Andover
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register