Skip to the content

FDA Medical Device Recalls (openFDA)

Cardiac Assist, Inc: Protek Solo Venous Dilator Set: dilator, vessel, for percutaneous catheterization. Catalog #: 5840-2417 - LS Tandem Life Kit - V24/A17 Catalog #: 5840-2415 - LS Tandem Life Kit - V24/A15 Catalog #: 5740-2417 - TandemLife Kit - V24/A17 Catalog #: 5100-1422 - ProtekSolo Venous Dilator Set (14-22 Fr)

Recall number
Z-1486-2021
Recalling firm
Cardiac Assist, Inc
Product
Protek Solo Venous Dilator Set: dilator, vessel, for percutaneous catheterization. Catalog #: 5840-2417 - LS Tandem Life Kit - V24/A17 Catalog #: 5840-2415 - LS Tandem Life Kit - V24/A15 Catalog #: 5740-2417 - TandemLife Kit - V24/A17 Catalog #: 5100-1422 - ProtekSolo Venous Dilator Set (14-22 Fr)
Status
Terminated
Initiated
2021-03-19
Posted
not on file
Terminated
2021-11-22
Reason
On 02/26/2021, it was found that there were incorrectly packaged Dilator sets in Finished Goods. On 03/1/2021, it was confirmed that 14 kits containing the impacted products had been distributed to the field.
Root cause
Under Investigation by firm
Action
On 3/11/2021 LivaNova issued an Urgent Medical Device Correction notice via letter notifying customers the firm identified an error with the assembly of a limited number of ProtekSolo Venous Dilator Sets. The ProtekSolo Venous Dilator Set 14-22 Fr should contain 3 dilators: 14Fr, 18Fr and 22Fr. A limited number of boxes labeled as ProtekSolo Venous Dilator Set 14-22 Fr incorrectly contain 4 dilators: 14Fr, 18Fr, 22Fr and 26Fr.
Quantity
13 kits
Distribution
US Nationwide distribution in the states of AL, AZ, IL, TN.
Lot, serial or model codes
LOT/SERIAL: 285586, 00145202, 00145204, 00871587, 00871549, 00871560, 00871561, 00871562, 00871563, 00871564, 00871565, 00871566, 00871567,   UDI:  (01) 00814112020463(17)231024(11)201024(10)285586 (01)00814112020517(17)220222(11)210118(21)00145202 (01)00814112020517(17)220222(11)210118(21)00145204 (01)00814112020708(17)210203(11)220117(21)00871587 (01)00814112020092(17)210129(11)220117(21)00871549 (01)00814112020092(17)210129(11)220117(21)00871560 (01)00814112020092(17)210129(11)220117(21)00871561 (01)00814112020092(17)210129(11)220117(21)00871562 (01)00814112020092(17)210129(11)220117(21)00871563 (01)00814112020092(17)210129(11)220117(21)00871564 (01)00814112020092(17)210129(11)220117(21)00871565 (01)00814112020092(17)210129(11)220117(21)00871566 (01)00814112020092(17)210129(11)220117(21)00871567
510(k) numbers
["K162181"]
PMA numbers
not on file
Product code
DRE
Firm FEI
3002228812
Firm city
Pittsburgh
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register