Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Healthcare Diagnostics, Inc: ADVIA Centaur CA 125II Assay; Cat No. 09427226, SMN 10310443 (1-pack/100 tests); Cat no. 01678114, SMN 10315696 (5-pack/500 tests)

Recall number
Z-1486-2018
Recalling firm
Siemens Healthcare Diagnostics, Inc
Product
ADVIA Centaur CA 125II Assay; Cat No. 09427226, SMN 10310443 (1-pack/100 tests); Cat no. 01678114, SMN 10315696 (5-pack/500 tests)
Status
Terminated
Initiated
2018-02-21
Posted
not on file
Terminated
2019-05-16
Reason
Upon dilution, some patient samples may exhibit over-recovery outside the representative data provided in the assay instructions for use.
Root cause
Other
Action
A Customer Notification was sent to customers on 2/21/18. The letter instructs customers to refer to the information provided until the appropriate updates are available in the assay instructions for use. The letter is to be retained and forwarded to those who may have received the product.
Quantity
SMN 10310443 (1-pack/100 tests) - 13257 kits; SMN 10315696 (5-pack/500 tests) - 9930 kits
Distribution
US Nationwide, and the following countries: Albania, Angola, Argentina, Armenia, Australia, Austria, Bahrain, Bangladesh, Belgium, Bosnia Herzegovina, Brazil, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Latvia, Lithuania, Luxembourg, Macedonia, Malaysia, Malta, Mexico, Morocco, Myamar, Nepal, Netherlands, New Zealand, Nicaragua, Norway, Pakistan, Paraguay, Peru, Philippines, Poland, Portugal, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan,Thailand, Turkey, United Arab Emirates, Ukraine, United Kingdom, Uruguay, Uzbekistan,Vatican City, and Vietnam.
Lot, serial or model codes
SMN 10310443 (1-pack/100 tests): Lot 71804174, Exp 2018-Apr-27, UDI (01)00630414461298(10)71804174(17)20180427 Lot 73838174, Exp 2018-Apr-27, UDI (01)00630414461298(10)73838174(17)20180427 Lot 88601174, Exp 2018-Apr-27, UDI (01)00630414461298(10)88601174(17)20180427 Lot 94549174, Exp 2018-Apr-27, UDI (01)00630414461298(10)94549174(17)20180427 Lot 23904176, Exp 2018-Aug-30, UDI (01)00630414461298(10)23904176(17)20180830 Lot 24439176, Exp 2018-Aug-30, UDI (01)00630414461298(10)24439176(17)20180830 Lot 31842176, Exp 2018-Aug-30, UDI (01)00630414461298(10)31842176(17)20180830 Lot 42051176, Exp 2018-Aug-30, UDI (01)00630414461298(10)42051176(17)20180830 Lot 54008176, Exp 2018-Aug-30, UDI (01)00630414461298(10)54008176(17)20180830  SMN 10315696 (5-pack/500 tests): Lot 71229174, Exp 2018-Apr-27, UDI (01)00630414461304(10)71229174(17)20180427 Lot 73455174, Exp 2018-Apr-27, UDI (01)00630414461304(10)73455174(17)20180427 Lot 84113174, Exp 2018-Apr-27, UDI (01)00630414461304(10)84113174(17)20180427 Lot 91568174, Exp 2018-Apr-27, UDI (01)00630414461304(10)91568174(17)20180427 Lot 23903176, Exp 2018-Aug-30, UDI (01)00630414461304(10)23903176(17)20180830 Lot 24944176, Exp 2018-Aug-30, UDI (01)00630414461304(10)24944176(17)20180830 Lot 41442176, Exp 2018-Aug-30, UDI (01)00630414461304(10)41442176(17)20180830 Lot 46735176, Exp 2018-Aug-30, UDI (01)00630414461304(10)46735176(17)20180830
510(k) numbers
["K020828"]
PMA numbers
not on file
Product code
LTK
Firm FEI
1219913
Firm city
East Walpole
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register