Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc: Siemens-branded linear accelerator with a 550 TxT Treatment Table with a serial number <_ 1894. Siemens brand ARTISTE, ONCOR and PRIMUS Linear Accelerator systems include the 550 TxT" Treatment Table. Manufactured by: SIEMENS AG, MEDICAL SOLUTIONS Roentgenstrasse 19-21 Kemnath, GERMANY 95478. Product Usage: The intended use of the SIEMENS branded ARTISTE, ONCOR and PRIMUS family of linear accelerator systems is to deliver X-ray photon and electron radiation for the therapeutic treatment of cancer.

Recall number
Z-1486-2012
Recalling firm
Siemens Medical Solutions USA, Inc
Product
Siemens-branded linear accelerator with a 550 TxT Treatment Table with a serial number <_ 1894. Siemens brand ARTISTE, ONCOR and PRIMUS Linear Accelerator systems include the 550 TxT" Treatment Table. Manufactured by: SIEMENS AG, MEDICAL SOLUTIONS Roentgenstrasse 19-21 Kemnath, GERMANY 95478. Product Usage: The intended use of the SIEMENS branded ARTISTE, ONCOR and PRIMUS family of linear accelerator systems is to deliver X-ray photon and electron radiation for the therapeutic treatment of cancer.
Status
Terminated
Initiated
2012-04-03
Posted
2012-05-03
Terminated
2013-03-14
Reason
SIEMENS has received reports where a dropping down of the 550 TxT" Treatment Table was observed while the table was being moved downwards during operation by means of the hand control.
Root cause
Software design
Action
Siemens sent Customer Information letters to all affected customers beginning April 3, 2012 via Siemens Field Service Engineers. The letter identified the affected product, what is the issue and what has Siemens done to address this issue. The letter included an Update Instructions TH006/12/S - TxT Table Software Update VD00A. SIEMENS has arranged for a modification of the affected system. A SIEMENS Service Representative will ensure that the new software version VD00A is installed on affected system. The letter instructs customer to please include this Customer Information in their Digital Linear Accelerator System Owner Manual chapter Safety Advisory Letters where it should remain.
Quantity
624 active devices
Distribution
Worldwide Distribution - USA (nationwide) including P.R and the countries of: Africa, Angola, Argentina, Australia, Austria, Belgium, Bosnia, Brazil, Bulgaria, Canada, Chile, China, Colombia, Croatia, Egypt, France, Germany, Greece, Hungary, India, Indonesia, Italy, Japan, Kuwait, Lebanon, Malaysia, Mexico, Myanmar, Netherlands, New Zealand, Norway, Pakistan, Philippines, Poland, Portugal, Republic of Korea, Russian Federation, Saudi Arabia, Serbia, South Thailand, Sri Lanka, Spain, Switzerland, Trinidad, Tobago, Turkey, Ukraine and United Kingdom, .
Lot, serial or model codes
System IVK no. Component Component IVK Ser #  ARTISTE MV system 8139789 550 TxT table 7346534 d1894 ONCOR Expression 7360717 550 TxT table 7346534 d1894 ONCOR Impression 5857912 550 TxT table 7346534 d1894 Plus ONCOR Impression 5857920 550 TxT table 7346534 d1894 ONCOR Avant Garde 5863472 550 TxT table 7346534 d1894 PRIMUS HI 4504200 550 TxT table 7346534 d1894 PRIMUS Mid-Energy 1940035 550 TxT table 7346534 d1894
510(k) numbers
["K060226","K103606"]
PMA numbers
not on file
Product code
IYE
Firm FEI
2937457
Firm city
Concord
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register