Skip to the content

FDA Medical Device Recalls (openFDA)

GE Healthcare: GE Precision RXi Digital Remote X-Ray Imaging R&F System. The Precision RXi digital is a remote R&F system consisting of a tilting table with integrated spot film device, 65 or 80 kW pulsing generator and high heat capacity X-Ray tube. The system can be configured to meet a wide variety of digital record and fluoroscopy exams for various clinical applications with either 32cm or 40cm image intensifier. GE Healthcare

Recall number
Z-1485-06
Recalling firm
GE Healthcare
Product
GE Precision RXi Digital Remote X-Ray Imaging R&F System. The Precision RXi digital is a remote R&F system consisting of a tilting table with integrated spot film device, 65 or 80 kW pulsing generator and high heat capacity X-Ray tube. The system can be configured to meet a wide variety of digital record and fluoroscopy exams for various clinical applications with either 32cm or 40cm image intensifier. GE Healthcare
Status
Terminated
Initiated
2006-11-01
Posted
2006-09-30
Terminated
2008-02-24
Reason
Automatic exposure control (AEC) automatically resets allowing another exposure without requiring manual reset as specified in 21 CFR 1020.31 (a)(3)(iv)
Root cause
Other
Action
Correction to devices is being implemented through GEHC Field Modification Instructions.
Quantity
83
Distribution
within US: CO, IL, LA, MA, MI, NC, PA, SC,TN, VA. OUS to include: Austrailia, Belgium, Burkina Faso, Brazil, Canada, China, Croatia, Czech Republic, Finland, France, Germany, Italy, Republic of Korea, Lebanon, Poland, Portugal, Russia, Slovakia, South Africa, Spain, Taiwan, Republic of China, United Kingdom
Lot, serial or model codes
198109, 229088, 241096, 117130, 156120, 116119, 127130, 174114, 193102, 183100, 252092, 261088, 148113, 149126, 184116, 165110, 188101, 164106, 157123, 23775004, 244092, 113127, 127114, 185099, 170109, 179118, 257083, 224091, 162108, 108130, 199100, 202104, 204100, 186106, 237092, 262080, 166119, 120116, 245094, 260086, 231092, 109129, 114127, 135125, 104125, 105126, 124130, 001000, 133131, 118128, 121135, 205104, 243091, 115126, 217100, 134128, 219096, 218096, 2338037, 122130, 110129, 140132, 23949001, 172112, 130130, 112131, 145124, 146119, 292081, 233089, 23778003, 211100, 173106, 283093, 213098, 230093, 107127, 225100, 291074, 176119, 264095, 144117, 255085
510(k) numbers
["K041605"]
PMA numbers
not on file
Product code
JAA
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register