Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Healthcare Diagnostics, Inc.: Atellica CH Urine Albumin (UAlb). Material Number: 11537225

Recall number
Z-1484-2026
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Product
Atellica CH Urine Albumin (UAlb). Material Number: 11537225
Status
Open, Classified
Initiated
2026-01-20
Posted
2026-03-02
Terminated
not on file
Reason
Falsely depressed UAlb patient results may occur. Affected samples with any urine albumin concentration above the UAlb measuring interval >40.0 mg/dL may report values between 6.2 mg/dL and 40.0 mg/dL.
Root cause
Under Investigation by firm
Action
Siemens Healthineers notified consignees on about 01/20/2026 via mailed FedEx letter titled Urgent Medical Device Correction. Consignees were informed that as temporary mitigation, for any sample with a UAlb result above 5.0 mg/dL (50 mg/L), reanalyze the sample with a x10 dilution. If the diluted result is consistent with the initial result, either result may be reported. If the diluted result is inconsistent with the initial result, the diluted result should be reported. If you prefer to automate this dilution, please refer to your internal laboratory procedures. Consignees were instructed to review the notification with their Medical Director to determine the appropriate course of action, including evaluation of any previously generated results, if applicable, and complete and return the provided Field Correction Effectiveness Check Form. A follow-up communication will be provided when Customer Actions are no longer required.
Quantity
4,885 units
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Brazil, Canada, Chile, Curacao,¿St¿Eus, Czech¿Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, Ireland, Israel, Italy, Korea, Latvia, Malaysia, Netherlands, New Zealand, Norway, Portugal, Romania, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, United¿Kingdom.
Lot, serial or model codes
Material Number: 11537225. UDI-DI: 00630414611099. Lot Numbers: All lot numbers.
510(k) numbers
not on file
PMA numbers
not on file
Product code
JIR
Firm FEI
2432235
Firm city
Tarrytown
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register