Skip to the content

FDA Medical Device Recalls (openFDA)

Wright Medical Technology Inc: VALOR(R), TARGETING GUIDE, REF 4150011000, Wright Medical Technology, Inc., 5677 Airline, Road, Arlington, TN 38002. The VALOR" Ankle Fusion Nail System is intended to facilitate tibiotalocalcaneal arthrodesis to treat severe foot/ankle deformity, arthritis, instability, and skeletal defects after tumor resection. These include Neuro-osteoarthropathy (Charcots Foot), Avascular Necrosis of the talus, failed joint replacement, failed ankle fusion, distal tibia fracture non-unions, Osteoarthritis, Rheumatoid Arthritis, and Pseudoarthrosis.

Recall number
Z-1484-2012
Recalling firm
Wright Medical Technology Inc
Product
VALOR(R), TARGETING GUIDE, REF 4150011000, Wright Medical Technology, Inc., 5677 Airline, Road, Arlington, TN 38002. The VALOR" Ankle Fusion Nail System is intended to facilitate tibiotalocalcaneal arthrodesis to treat severe foot/ankle deformity, arthritis, instability, and skeletal defects after tumor resection. These include Neuro-osteoarthropathy (Charcots Foot), Avascular Necrosis of the talus, failed joint replacement, failed ankle fusion, distal tibia fracture non-unions, Osteoarthritis, Rheumatoid Arthritis, and Pseudoarthrosis.
Status
Terminated
Initiated
2012-04-05
Posted
2012-05-02
Terminated
2013-01-28
Reason
The possibility exists that the handle of the targeting guide may rotate and/or break under normal usage.
Root cause
Device Design
Action
Wright Medical Technology, Inc. sent an "URGENT PRODUCT RECALL NOTIFICATION" letter dated April 2, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter instructs customers to check their stock and to return the affected product to the firm. Additionally, a Product Recall Verification Form was enclosed for customers to complete and return via fax to 901-867-7401. Contact the firm at 901-867-4324 for questions regarding this notice.
Quantity
133 units
Distribution
Worldwide Distribution-USA (nationwide) and the country of Australia.
Lot, serial or model codes
all lots distributed until 03/9/2012
510(k) numbers
["K090857"]
PMA numbers
not on file
Product code
HSB
Firm FEI
1043534
Firm city
Arlington
Firm state
TN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register