FDA Medical Device Recalls (openFDA)
Wright Medical Technology Inc: VALOR(R), TARGETING GUIDE, REF 4150011000, Wright Medical Technology, Inc., 5677 Airline, Road, Arlington, TN 38002. The VALOR" Ankle Fusion Nail System is intended to facilitate tibiotalocalcaneal arthrodesis to treat severe foot/ankle deformity, arthritis, instability, and skeletal defects after tumor resection. These include Neuro-osteoarthropathy (Charcots Foot), Avascular Necrosis of the talus, failed joint replacement, failed ankle fusion, distal tibia fracture non-unions, Osteoarthritis, Rheumatoid Arthritis, and Pseudoarthrosis.
- Recall number
- Z-1484-2012
- Recalling firm
- Wright Medical Technology Inc
- Product
- VALOR(R), TARGETING GUIDE, REF 4150011000, Wright Medical Technology, Inc., 5677 Airline, Road, Arlington, TN 38002. The VALOR" Ankle Fusion Nail System is intended to facilitate tibiotalocalcaneal arthrodesis to treat severe foot/ankle deformity, arthritis, instability, and skeletal defects after tumor resection. These include Neuro-osteoarthropathy (Charcots Foot), Avascular Necrosis of the talus, failed joint replacement, failed ankle fusion, distal tibia fracture non-unions, Osteoarthritis, Rheumatoid Arthritis, and Pseudoarthrosis.
- Status
- Terminated
- Initiated
- 2012-04-05
- Posted
- 2012-05-02
- Terminated
- 2013-01-28
- Reason
- The possibility exists that the handle of the targeting guide may rotate and/or break under normal usage.
- Root cause
- Device Design
- Action
- Wright Medical Technology, Inc. sent an "URGENT PRODUCT RECALL NOTIFICATION" letter dated April 2, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter instructs customers to check their stock and to return the affected product to the firm. Additionally, a Product Recall Verification Form was enclosed for customers to complete and return via fax to 901-867-7401. Contact the firm at 901-867-4324 for questions regarding this notice.
- Quantity
- 133 units
- Distribution
- Worldwide Distribution-USA (nationwide) and the country of Australia.
- Lot, serial or model codes
- all lots distributed until 03/9/2012
- 510(k) numbers
- ["K090857"]
- PMA numbers
- not on file
- Product code
- HSB
- Firm FEI
- 1043534
- Firm city
- Arlington
- Firm state
- TN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28