FDA Medical Device Recalls (openFDA)
B Braun Medical Inc: Brand Name: B. Braun Product Name: 21GA WINGED INF SET Model/Catalog Number: 7M2802 Software Version: N/A Product Description: A device designed for subcutaneous implantation and intended to be used temporarily to expand an implanted tissue expander by percutaneous injection of air or fluids. Component: N/A
- Recall number
- Z-1483-2026
- Recalling firm
- B Braun Medical Inc
- Product
- Brand Name: B. Braun Product Name: 21GA WINGED INF SET Model/Catalog Number: 7M2802 Software Version: N/A Product Description: A device designed for subcutaneous implantation and intended to be used temporarily to expand an implanted tissue expander by percutaneous injection of air or fluids. Component: N/A
- Status
- Open, Classified
- Initiated
- 2026-02-02
- Posted
- 2026-03-02
- Terminated
- not on file
- Reason
- The potential for the needle tip to be dull/blunt, difficult to advance, and/or break.
- Root cause
- Process control
- Action
- On February 2, 2026 URGENT MEDICAL DEVICE RECALL NOTIFICATION letters were sent to customers. Actions to be taken: 1. Review the notice in its entirety and ensure that all users in your organization are aware and informed about this voluntary recall. If you are a distributor and have further distributed the product, please forward this notice to your consignees. The recall is to be extended to the hospital/healthcare facility level. 2. Examine inventory for affected product, cease use and quarantine product subject to recall. Do not destroy any affected product. 5. Once we receive your Acknowledgement Form, a B. Braun Customer Support representative will contact you with instructions on how to return any impacted cases, including partial cases, in your possession and provide credit and/or replacement of the product based on your individual needs.
- Quantity
- 51300 units
- Distribution
- Worldwide - US Nationwide distribution in the state of TX and the country of Costa Rica.
- Lot, serial or model codes
- Model/Catalog Number: 7M2802; Primary-DI 4046964787102, Unit of Use- DI 4046964349904; Lot 0062003864, Exp 30APR2032; Lot 0062004133, Exp 30JUN2032; Lot 0062006454, Exp 31JUL2032; Lot 0062028929, Exp 31AUG2032; Lot 0062037141, Exp 30SEP2032, Lot 0062037691, Exp 30SEP2032.
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- LCJ
- Firm FEI
- 2521402
- Firm city
- Bethlehem
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28