FDA Medical Device Recalls (openFDA)
Aesculap Implant Systems LLC: COLUMBUS REV F TIBIA OFFSET CEMENTED, Knee implant components, various sizes, model nos. NR068K NR068Z NR070K NR070Z NR071K NR071Z NR072K NR072Z NR073K NR073Z NR074K NR074Z NR075K NR075Z NR076K NR076Z NR077K NR077Z NR078K NR078Z NR079K NR079Z Product Usage: Tibia-Plateaus are part of implants in the group of knee endoprosthetics. For these implant parts there are cemented and cement-free versions as well as versions with zirconium nitride multi layer.
- Recall number
- Z-1482-2020
- Recalling firm
- Aesculap Implant Systems LLC
- Product
- COLUMBUS REV F TIBIA OFFSET CEMENTED, Knee implant components, various sizes, model nos. NR068K NR068Z NR070K NR070Z NR071K NR071Z NR072K NR072Z NR073K NR073Z NR074K NR074Z NR075K NR075Z NR076K NR076Z NR077K NR077Z NR078K NR078Z NR079K NR079Z Product Usage: Tibia-Plateaus are part of implants in the group of knee endoprosthetics. For these implant parts there are cemented and cement-free versions as well as versions with zirconium nitride multi layer.
- Status
- Terminated
- Initiated
- 2020-01-21
- Posted
- not on file
- Terminated
- 2021-09-22
- Reason
- Low Density Polyethylene (LDPE) particles may be present on the surface of the implant upon the opening of the inner packaging.
- Root cause
- Material/Component Contamination
- Action
- On January 21, 2020, the firm distributed Urgent Medical Device Recall Notification letters to affected customers. Customers were informed that LDPE particles may be present on the surface of implants upon the opening of the inner packaging. The firm has developed a new packaging configuration to mitigate the issue. Until the product in the field is removed and replaced with product in the new packaging configuration, the firm has taken immediate action to limit the transportation cycles of the product to six months. Customers were asked to immediately examine their inventory and identify product subject to recall. In addition, if customers have further distributed the product, please identify customers and notify them at once of the recall. Customers should NOT destroy affected product. Customers were asked to complete the customer response form and indicate the total number of units in their possession. A representative of Aesculap will reach out to customer facilities to schedule site visits for removal of affected product and replacement of inventory.
- Quantity
- not on file
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of AL, CA, CO, FL, GA, ID, IL, IN, KS, LA, MA, MD, MI, MN, MO, NC, OH, OK, OR, PA, TN, TX, UT, WA and countries of Saudi Arabia, Utd. Arab Emir., Argentina, Austria, Australia, Bosnia-Herz., Bulgaria, Brazil, Switzerland, Chile, China, Colombia, Costa Rica, Cuba, Czech Republic, Germany, Algeria, Ecuador, Estonia, Spain, United Kingdom, Georgia, Greece, Hong Kong, Croatia, Hungary, Indonesia, Ireland, Israel, India, Iraq, Iran, Italy, Jordan, Japan, South Korea, Lithuania, Latvia, Libya, Moldavia, Mexico, Malaysia, Nigeria, Netherlands, Poland, Portugal, Paraguay, Russian Fed., Singapore, El Salvador, Thailand, Turkey, Ukraine, South Africa.
- Lot, serial or model codes
- All product with old packaging configuration. New packaging configuration begins with Lot 52533610.
- 510(k) numbers
- ["K083772"]
- PMA numbers
- not on file
- Product code
- JWH
- Firm FEI
- 2916714
- Firm city
- Center Valley
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28