Aesculap Implant Systems LLC: PRESTIGE ATRAUMATIC GRASPER W/PORT DUAL ACTION, ROTATING 5mm DIAM., 36cm LENGTH. Product Code 8360-10

Recall number
Z-1481-2022
Recalling firm
Aesculap Implant Systems LLC
Product
PRESTIGE ATRAUMATIC GRASPER W/PORT DUAL ACTION, ROTATING 5mm DIAM., 36cm LENGTH. Product Code 8360-10
Status
Completed
Initiated
2020-04-01
Posted
not on file
Terminated
not on file
Reason
Endoscopic graspers may become separated at the weld from handle to the shaft of the device.
Root cause
Process change control
Action
A notification letter was issued April 30, 2020 sent USPS Certified mail . The notification identifies the products as Catalog Item # 8360-10 PRESTIGE ATRA GRASPER DBL-ACT 5MM and 8361-10 PRESTIGE RETR GRASPER DBL-ACT 5MM. Customers were to evaluate inventory with a "tug test" and submit the notification acknowledgment form. Should a device fail the tug test, customers were to file a complaint using Aesculap's Online Customer Complaint Submission Form and return a the product. The products have since been discontinued.
Quantity
3392
Distribution
United States Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY, Washington DC and Puerto Rico.
Lot, serial or model codes
UDI-DI: 04046955083374, lots M45041; M45399; M45711; M46069; M46350; M47462; M47893; M48115; M48279; M48346; M48467; M48596; M48753; M48848; M48907; M49375; M49435; M49632; M49683; M49711; M49867; M55796.
510(k) numbers
not on file
PMA numbers
not on file
Product code
KOA
Firm FEI
2916714
Firm city
Center Valley
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register