Skip to the content

FDA Medical Device Recalls (openFDA)

Staar Surgical AG: EVO+ Visian Toric Implantable Collamer Lens (TICL), Model: VTICM5

Recall number
Z-1480-2026
Recalling firm
Staar Surgical AG
Product
EVO+ Visian Toric Implantable Collamer Lens (TICL), Model: VTICM5
Status
Open, Classified
Initiated
2026-01-21
Posted
2026-03-02
Terminated
not on file
Reason
Toric Implantable Collamer for correction/reduction of myopia in phakic eyes with astigmatism were labeled as 12.1 mm length but are 12.6 mm length lenses, so there is the potential for excessive vault, which may cause: Significant reduction of irido-corneal angles, Significant shallowing of anterior chamber, Patient discomfort, Elevated IOP/Angle closure, Unreactive pupil, Lens exchange/removal.
Root cause
Process control
Action
On 1/21/2026, a Field Safety Notice was mailed to customers who were asked to do the following: 1. If any of the devices identified in the Field Safety Notice Return Response form have not already been implanted, remove them from your inventory and quarantine the product. 2. Complete the Field Safety Notice Return Response form indicating if each device has been implanted or is in your possession. 3. Complete and email the Field Safety Notice Return Response form to customerservice.ag@staar.com. 5. Return affected devices to the firm. Should you have any questions, contact the firm's Representative or Customer Service at customerservice.ag@staar.com or by phone at +41323328888 Monday through Friday, 8:00am to 5:00pm CET.
Quantity
7
Distribution
International distribution to the countries of India, Iran, Korea, Saudi Arabia.
Lot, serial or model codes
UDI-DI: 00840311304363. Expiration Date: 2027-11-30 Serial Numbers: T2286964, T2286971, T2286973, T2286979, T2286980, T2286981, T2286988
510(k) numbers
not on file
PMA numbers
["P030016"]
Product code
QCB
Firm FEI
3002808259
Firm city
Nidau
Firm state
not on file
Firm country
Switzerland

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register