Skip to the content

FDA Medical Device Recalls (openFDA)

Aesculap Implant Systems LLC: COLUMBUS CRA/PSA TIB.PLAT.CEMENTED, Knee implant components, various sizes, model nos. NN469K NN470K NN471K NN472K NN473K NN474K NN475K NN476K NN477K NN478K NN479K Product Usage: Tibia-Plateaus are part of implants in the group of knee endoprosthetics. For these implant parts there are cemented and cement-free versions as well as versions with zirconium nitride multi layer.

Recall number
Z-1480-2020
Recalling firm
Aesculap Implant Systems LLC
Product
COLUMBUS CRA/PSA TIB.PLAT.CEMENTED, Knee implant components, various sizes, model nos. NN469K NN470K NN471K NN472K NN473K NN474K NN475K NN476K NN477K NN478K NN479K Product Usage: Tibia-Plateaus are part of implants in the group of knee endoprosthetics. For these implant parts there are cemented and cement-free versions as well as versions with zirconium nitride multi layer.
Status
Terminated
Initiated
2020-01-21
Posted
not on file
Terminated
2021-09-22
Reason
Low Density Polyethylene (LDPE) particles may be present on the surface of the implant upon the opening of the inner packaging.
Root cause
Material/Component Contamination
Action
On January 21, 2020, the firm distributed Urgent Medical Device Recall Notification letters to affected customers. Customers were informed that LDPE particles may be present on the surface of implants upon the opening of the inner packaging. The firm has developed a new packaging configuration to mitigate the issue. Until the product in the field is removed and replaced with product in the new packaging configuration, the firm has taken immediate action to limit the transportation cycles of the product to six months. Customers were asked to immediately examine their inventory and identify product subject to recall. In addition, if customers have further distributed the product, please identify customers and notify them at once of the recall. Customers should NOT destroy affected product. Customers were asked to complete the customer response form and indicate the total number of units in their possession. A representative of Aesculap will reach out to customer facilities to schedule site visits for removal of affected product and replacement of inventory.
Quantity
not on file
Distribution
Worldwide distribution - US Nationwide distribution in the states of AL, CA, CO, FL, GA, ID, IL, IN, KS, LA, MA, MD, MI, MN, MO, NC, OH, OK, OR, PA, TN, TX, UT, WA and countries of Saudi Arabia, Utd. Arab Emir., Argentina, Austria, Australia, Bosnia-Herz., Bulgaria, Brazil, Switzerland, Chile, China, Colombia, Costa Rica, Cuba, Czech Republic, Germany, Algeria, Ecuador, Estonia, Spain, United Kingdom, Georgia, Greece, Hong Kong, Croatia, Hungary, Indonesia, Ireland, Israel, India, Iraq, Iran, Italy, Jordan, Japan, South Korea, Lithuania, Latvia, Libya, Moldavia, Mexico, Malaysia, Nigeria, Netherlands, Poland, Portugal, Paraguay, Russian Fed., Singapore, El Salvador, Thailand, Turkey, Ukraine, South Africa.
Lot, serial or model codes
All product with old packaging configuration. New packaging configuration begins with Lot 52533610.
510(k) numbers
["K053390"]
PMA numbers
not on file
Product code
JWH
Firm FEI
2916714
Firm city
Center Valley
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28