FDA Medical Device Recalls (openFDA)
Captiva Spine, Inc: Captiva's lumbar intervertebral fusion system Reamers (color code rings on the Reamers).
- Recall number
- Z-1480-2015
- Recalling firm
- Captiva Spine, Inc
- Product
- Captiva's lumbar intervertebral fusion system Reamers (color code rings on the Reamers).
- Status
- Terminated
- Initiated
- 2015-02-25
- Posted
- 2015-04-21
- Terminated
- 2015-06-01
- Reason
- It is possible for the titanium color-coded ring to detach from the Reamers.
- Root cause
- Device Design
- Action
- On 02/26/2015 the firm sent a Voluntary Correction Notification to its consignees via FedEx. Within that notification the firm asks to please check with your product representative and Central Supply Departments to determine if you have any of the products listed on the notificaiton. If any product is located please call 877-772-5571 from 8:30am (Eastern Time) to receive instructions as to how to remove the color-coded ring OR please follow the enclosed steps for the correction in their entirety. They contain information about returning the removed rings. Enclosed is Form 30.4 Product Tracking and Verification Form. You must complete and return this form even if you do not have any rings to remove and return. Please contact your Captiva Spine customer service representative if you have any questions regarding this correction, any of our products, or would like assistance with the correction.
- Quantity
- 233 devices
- Distribution
- GA, IN, FL, UT, AZ, CA, and NY
- Lot, serial or model codes
- Part #: PI-0507, Lot #s:1308811A, 1457704A; Part #: PI-0508, Lot #s: 1308812A, 1457705A; Part #: PI-0509, Lot #s: 1038813A, 1457706A; Part #: PI-0510, Lot #s: 1308814A, 1457707A; Part #: PI-0511, Lot #s: 1308815A, 1457708A; Part #: PI-0512, Lot #s: 1308816A, 1457709A; Part #: PI-0513, Lot #s: 1308817A, 1457710A; Part #: PI-0514, Lot #s: 1308818A, 1457711A; Part #: PI-0507-S, Lot #: 2130005; Part #: PI-0508-S, Lot #: 2130006; Part #: PI-0509-S, Lot #: 2130007; Part #: PI-0510-S, Lot #: 2130008; Part #: PI-0511-S, Lot #: 2130009; Part #: PI-0512-S, Lot #: 2130010; Part #: PI-0513-S, Lot #:2130011; Part #: PI-0514-S, Lot #: 2130012
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- LXH
- Firm FEI
- 3006082533
- Firm city
- Jupiter
- Firm state
- FL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28