Skip to the content

FDA Medical Device Recalls (openFDA)

Bio-Rad Laboratories Inc: Bio Rad brand Autoimmune EIA ANA Screening Kit, 96 Tests, Catalog No. 96AN-BPU-TRINITY. Manufactured by: Bio-Rad Laboratories, Inc. Hercules, CA and distributed by Trinity Biotech, USA. Intended to screen for the presence of antinuclear antibodies (ANAs) in human serum as an aid in the diagnosis of certain rheumatic diseases.

Recall number
Z-1479-2011
Recalling firm
Bio-Rad Laboratories Inc
Product
Bio Rad brand Autoimmune EIA ANA Screening Kit, 96 Tests, Catalog No. 96AN-BPU-TRINITY. Manufactured by: Bio-Rad Laboratories, Inc. Hercules, CA and distributed by Trinity Biotech, USA. Intended to screen for the presence of antinuclear antibodies (ANAs) in human serum as an aid in the diagnosis of certain rheumatic diseases.
Status
Terminated
Initiated
2010-03-01
Posted
2011-03-02
Terminated
2011-03-11
Reason
The firm received customer complaints of increased positivity rates associated with the ANA Screening kit due to a decline in stability.
Root cause
Nonconforming Material/Component
Action
An Urgent: Medical Device Recall Notification, dated March 1, 2010, was provided to the firm's consignee, identifying the affected product and stating the reason for the recall. The letter also asked the customer to examine their inventory and quarantine the affected kits. If the product had been further distributed, those customers must be identified, notified, and asked to discontinue use of the affected product. Customers were to discard all affected product and complete and return the enclosed Medical Device Recall Response Form. The letter also stated that replacement kits would be available in mid-March.
Quantity
2,109 bulk packs
Distribution
Nationwide Distribution -- New York, US.
Lot, serial or model codes
Lot Numbers: 24591524, exp 10/31/2010; 24591936, exp 12/31/2010; and 24500022, exp 12/31/2010.
510(k) numbers
["K954723"]
PMA numbers
not on file
Product code
LKJ
Firm FEI
2915274
Firm city
Hercules
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register