FDA Medical Device Recalls (openFDA)
Abbott Spine: Sequoia Spinal System Surgical Kits; Pedicle screw spinal system; Non-sterile (kits intended to be sterilized prior to use); Rx Only; The Sequoia Spinal System is designed to aid in the surgical correction of several types of spinal conditions. The product is a system of smaller devices used to provide stabilization during the development of a solid fusion mass.
- Recall number
- Z-1479-2008
- Recalling firm
- Abbott Spine
- Product
- Sequoia Spinal System Surgical Kits; Pedicle screw spinal system; Non-sterile (kits intended to be sterilized prior to use); Rx Only; The Sequoia Spinal System is designed to aid in the surgical correction of several types of spinal conditions. The product is a system of smaller devices used to provide stabilization during the development of a solid fusion mass.
- Status
- Terminated
- Initiated
- 2008-02-13
- Posted
- 2008-07-23
- Terminated
- 2008-07-28
- Reason
- Screw Head Dissociation: The polyaxial screw head may dissociate from screw shaft during surgery. Also, the polyaxial screw driver is experiencing difficulty (broken/bent driver shafts, difficulty attaching screws to the driver and jamming of the collet.)
- Root cause
- Device Design
- Action
- The recalling firm notified sales reps regarding the recall by phone on 02/13/08. The sales reps were instructed to discontinue kit use in surgeries and to return inventory. Surgeons who had used the Sequoia System in surgery were notified of the recall via letter on 02/20/08. Surgeons were asked to monitor for post-operative dissociation in implanted patients and to notify Abbott Spine sales reps if any incidences occur.
- Quantity
- 38 surgical kits.
- Distribution
- Nationwide Distribution including states of AR, AZ, CA, CT, CO, FL, IL, MD, MI, MN, NC, OH, OK, OR, PA, TX, VA and WI.
- Lot, serial or model codes
- Part Numbers: 3300-0001-PL and 3300-0002-PL.
- 510(k) numbers
- ["K072672"]
- PMA numbers
- not on file
- Product code
- NKB
- Firm FEI
- 3000718496
- Firm city
- Austin
- Firm state
- TX
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28