Abbott Spine: Sequoia Spinal System Surgical Kits; Pedicle screw spinal system; Non-sterile (kits intended to be sterilized prior to use); Rx Only; The Sequoia Spinal System is designed to aid in the surgical correction of several types of spinal conditions. The product is a system of smaller devices used to provide stabilization during the development of a solid fusion mass.

Recall number
Z-1479-2008
Recalling firm
Abbott Spine
Product
Sequoia Spinal System Surgical Kits; Pedicle screw spinal system; Non-sterile (kits intended to be sterilized prior to use); Rx Only; The Sequoia Spinal System is designed to aid in the surgical correction of several types of spinal conditions. The product is a system of smaller devices used to provide stabilization during the development of a solid fusion mass.
Status
Terminated
Initiated
2008-02-13
Posted
2008-07-23
Terminated
2008-07-28
Reason
Screw Head Dissociation: The polyaxial screw head may dissociate from screw shaft during surgery. Also, the polyaxial screw driver is experiencing difficulty (broken/bent driver shafts, difficulty attaching screws to the driver and jamming of the collet.)
Root cause
Device Design
Action
The recalling firm notified sales reps regarding the recall by phone on 02/13/08. The sales reps were instructed to discontinue kit use in surgeries and to return inventory. Surgeons who had used the Sequoia System in surgery were notified of the recall via letter on 02/20/08. Surgeons were asked to monitor for post-operative dissociation in implanted patients and to notify Abbott Spine sales reps if any incidences occur.
Quantity
38 surgical kits.
Distribution
Nationwide Distribution including states of AR, AZ, CA, CT, CO, FL, IL, MD, MI, MN, NC, OH, OK, OR, PA, TX, VA and WI.
Lot, serial or model codes
Part Numbers: 3300-0001-PL and 3300-0002-PL.
510(k) numbers
["K072672"]
PMA numbers
not on file
Product code
NKB
Firm FEI
3000718496
Firm city
Austin
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register