Stryker Instruments: Stryker 2.3mm Tapered Router, Catalog no. 5820-071-023 instrument used in craniotomy procedures.

Recall number
Z-1478-2022
Recalling firm
Stryker Instruments
Product
Stryker 2.3mm Tapered Router, Catalog no. 5820-071-023 instrument used in craniotomy procedures.
Status
Open, Classified
Initiated
2022-06-24
Posted
not on file
Terminated
not on file
Reason
There is a potential for the core diameter to be undersized, which may lead to the router breaking. A break can lead to unintended metal fragments in the surgical site, resulting in the potential for surgical intervention or damage to critical neurological or vascular structures.
Root cause
Process change control
Action
On June 24, 2022, the firm notified customers via Urgent Medical Device Recall Letters. Customers were instructed to locate and quarantine any affected products found in their inventory. Upon receiving the customer's completed response form, the firm will provide a shipping label to return any recalled product. Upon receipt of the recalled product, Stryker will contact the customer to arrange for replacement product.
Quantity
543
Distribution
Worldwide distribution - US Nationwide distribution in the states of Arizona, California, Florida, Georgia, Iowa, Michigan, Missouri, New Mexico, North Carolina, Ohio, Texas, Virginia and the countries of Canada, Netherlands, and Thailand.
Lot, serial or model codes
UDI-DI (GTIN): 07613327294910 Affected Lots: 22011017 21330017 20216017 20140017 20139027
510(k) numbers
["K143320"]
PMA numbers
not on file
Product code
HBE
Firm FEI
3015967359
Firm city
Portage
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register