Terumo Cardiovascular Systems Corporation: Flowmeter Module (accessory to Terumo Advanced Perfusion System 1). Provides the interface between the flow sensor and the system.

Recall number
Z-1478-2018
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
Flowmeter Module (accessory to Terumo Advanced Perfusion System 1). Provides the interface between the flow sensor and the system.
Status
Terminated
Initiated
2018-01-19
Posted
not on file
Terminated
2018-07-18
Reason
Inaccurrate flow readings. Depending on the degree of inaccuracy, this issue may not be easy for the user to detect during setup or use (for example, following a Flow Probe relocation or manipulation).
Root cause
Employee error
Action
The firm notified customers via phone alerting them of this affected device, issue, potential hazard, correction, and instructions. The communication also included scheduling an expedited service call for a Field Service Technician to replace the affected Flowmeter Module with a corrected Flowmeter Module. When necessary to avoid delaying or cancelling life-sustaining surgery, users can continue to use the Flowmeter Module while awaiting replacement. Once the issue is recognized by the user, if a replacement Flowmeter Module is available, replacement and reassignment of safety connections of the Flowmeter Module can be accomplished in less than 15 seconds. In the event that a replacement Flowmeter Module is not available, a less common mitigation is the use of a back-up stand-alone centrifugal pump or a stand-alone ultrasonic flowmeter system to provide flow data. Customers should receive a copy of the Urgent Medical Device Recall phone script by e-mail along with a Customer Response Form. The e-mail should be reviewed, and the form completed and returned as indicated. Questions or concerns can be directed Terumo CVS Customer Service at 1-800-521-2818.
Quantity
7
Distribution
Worldwide Distribution -- USA, including the states of IL, KY, and MI; and, the country of Singapore.
Lot, serial or model codes
Catalog # - 802018 Serial #: 01662, 01672, 01680, 01681,01682, 01683 , and 01687. UDI: 00886799000687.
510(k) numbers
["K172220"]
PMA numbers
not on file
Product code
DTQ
Firm FEI
1828100
Firm city
Ann Arbor
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register