Skip to the content

FDA Medical Device Recalls (openFDA)

BioMerieux SA: API Listeria (REF 10300). ZYM B reagent is used and included inside of API NH (REF 10400) and API Listeria (10300). All of the APIs are standardized systems combining several biochemical tests that enable group or species identification of most microorganisms.

Recall number
Z-1478-2014
Recalling firm
BioMerieux SA
Product
API Listeria (REF 10300). ZYM B reagent is used and included inside of API NH (REF 10400) and API Listeria (10300). All of the APIs are standardized systems combining several biochemical tests that enable group or species identification of most microorganisms.
Status
Terminated
Initiated
2014-03-06
Posted
2014-04-18
Terminated
2017-03-14
Reason
bioMerieux identified a visual defect and activity issue on the ZYM B reagent (REF 70493) and FB reagent (REF 70562). The activity issue lead to a false negative result of some Rapid ID 32 strips and API biochemical tests using the ZYM B reagent or FB reagent
Root cause
Other
Action
Letters were issued to customers on March 10, 2014. The letter contained instructions to be used if they still were using the product. The letter also included a report form which was to be returned.
Quantity
33,853
Distribution
Worldwide Distribution-USA (nationwide) including DC and Puerto Rico, and the states of AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, VA, WV, WA, WI, and WY and the country of Algeria, Angola, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Benin, Bolivia, Bosnia, Brazil, Brunel, Bulgaria, Burkina Faso, Cambodia, Cameroon, Canada, Central African Republic, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Dutch Antilles, Ecuador, Egypt, El Salvador, Estonia, Erythrea, France, Gabon, Georgia, Germany, Ghana, Greece, Guatemala, Haiti, Herzegovina, Honduras, Hungary, India, Indonesia, Iraq, Israel, Italy, Japan, Jamaica, Jordan, Kazakhstan, Kenya, Korea, Kosovo, Kuwait, Latvia, Lebanon, Lithuania, Macedonia, Malaysia, Mali, Mexico, Moldavia, Morocco, Mozambique, Myanmar, Netherlands, Nigeria, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Russia, Saudi Arabia, Senegal, Serbia, Singapore, Slovenia, South Africa, Spain, Sweden, Thailand, Trinidad & Tobago, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, and Vietnam.
Lot, serial or model codes
Code - 1002569040, Exp. Date 8/29/2014, 1002569041, Exp. Date 8/296/2014, 1002587190, Exp. Date 9/3/2014, 1002856790, Exp. Date 7/18/2014, 1002856800, Exp. Date 6/12/2014
510(k) numbers
not on file
PMA numbers
not on file
Product code
LIB
Firm FEI
1000611339
Firm city
Marcy L'Etoile
Firm state
not on file
Firm country
France

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register