Skip to the content

FDA Medical Device Recalls (openFDA)

C2 Therapeutics, Inc.: C2 CryoBalloon Ablation System - Controller Cap. C2 CryoBalloon Controller; Cryosurgical Unit, Cryogenic Surgical Device The affected product is the Controller Cap component only, which is provided to the customer already assembled onto the Controller. Model number: FG-1012 Product Usage: The C2 CryoBalloon Ablation System is intended for use as a cryosurgical tool in the field of general surgery, specifically for endoscopic applications, to include ablation of Barrett s Esophagus with dysplasia. It is a cryosurgical system with a nitrous oxide cooled balloon that is compatible with a commercially available endoscope with a minimum working channel inner diameter (ID) of 3.7 mm and length of 100 cm. The C2 CryoBalloon System is a system comprised of a Catheter (sterile), Controller (non-sterile), and Cartridge (non-sterile). The Device is used to ablate unwanted tissue by application of extreme cold. The balloon at the distal end of the Catheter comes in contact with tissue and is inflated with nitrous oxide. Tissue is visualized through the inflated balloon, and the treatment site is selected by adjusting the endoscope and Controller position. The nitrous oxide spray cools the balloon to ablate the unwanted tissue, and the nitrous oxide exhausts through the Controller.

Recall number
Z-1477-2017
Recalling firm
C2 Therapeutics, Inc.
Product
C2 CryoBalloon Ablation System - Controller Cap. C2 CryoBalloon Controller; Cryosurgical Unit, Cryogenic Surgical Device The affected product is the Controller Cap component only, which is provided to the customer already assembled onto the Controller. Model number: FG-1012 Product Usage: The C2 CryoBalloon Ablation System is intended for use as a cryosurgical tool in the field of general surgery, specifically for endoscopic applications, to include ablation of Barrett s Esophagus with dysplasia. It is a cryosurgical system with a nitrous oxide cooled balloon that is compatible with a commercially available endoscope with a minimum working channel inner diameter (ID) of 3.7 mm and length of 100 cm. The C2 CryoBalloon System is a system comprised of a Catheter (sterile), Controller (non-sterile), and Cartridge (non-sterile). The Device is used to ablate unwanted tissue by application of extreme cold. The balloon at the distal end of the Catheter comes in contact with tissue and is inflated with nitrous oxide. Tissue is visualized through the inflated balloon, and the treatment site is selected by adjusting the endoscope and Controller position. The nitrous oxide spray cools the balloon to ablate the unwanted tissue, and the nitrous oxide exhausts through the Controller.
Status
Terminated
Initiated
2017-01-06
Posted
2017-03-15
Terminated
2017-09-13
Reason
The Controller Cap of the C2 CyroBalloon Ablation System may crack as it is tightened onto the controller. That could compromise the integrity and result in damage to the Controller Cap.
Root cause
Under Investigation by firm
Action
Firm initially notified consignees who would be using the device in an upcoming procedure on January 6, 2017 to alert them of the potential issue and recommended that products with the affected cap not be used unless it can be checked by a C2 rep. Controller caps will be replaced within a few weeks; facilities will receive replacement caps by mail along with a shipping inventory label for return of those currently on hand. Formal recall letters were sent on January 19, 2017. Letters were sent by e-mail with receipt and read confirmation and follow up by phone call.
Quantity
755 controller caps
Distribution
Worldwide Distribution - US Nationwide and the countries of Germany, Belgium, The Netherlands United Kingdom and China.
Lot, serial or model codes
All lots distributed between June 23, 2016 and December 20, 2016: UPDATE - expanded to include product distributed from February 1 2016 through February 8, 2017. All numbers below have been updated. Lot numbers: Expiry dates 01122016-02 12-Jul-16; 06242016-01 1-Jan-2017; 11282016-03 2-Jun-2017; 01042016-03 14-Jul-16; 06292016-02 6-Jan-2017; 09202016-01 21-Mar-2017; 07062016-01 8-Jan-2017; 09222016-01 23-Mar-2017; 01252016-02 02-Aug-16; 07082016-02 13-Jan 2017; 09232016-01 28-Mar-2017; 01282016-01 01-Aug-16; 07082016-03 14-Jan-2017; 09232016-02 27-Mar-2017; 01292016-01 02-Aug-16; 07152016-01 21-Jan-2017; 09292016-01 4-Apr-2017; 02022016-03 09-Aug-16; 07152016-02 20-Jan-2017; 09302016-03 4-Apr-2017; 02082016-01 09-Aug-16; 07212016-01 25-Jan-2017; 10052016-02 19-Apr-2017; 02092016-02 11-Aug-16; 07252016-01 28-Jan-2017; 10142016-01 18-Apr-2017; 02112016-02 17-Aug-16; 07262016-01 9-Feb-2017; 10192016-01 24-Apr-2017; 02172016-02 25-Aug-16; 07282016-02 2-Feb-2017; 10212016-01 25-Apr-2017; 02222016-01 26-Aug-16; 08022016-01 17-Feb-2017; 10242016-01 26-Apr-2017; 03012016-01 04-Sep-16; 08032016-02 5-Feb-2017; 10262016-02 1-May-2017; 03072016-01 10-Sep-16; 08112016-01 16-Feb-2017; 10282016-02 8-May-2017; 03102016-02 15-Sep-16; 08232016-01 26-Feb-2017; 11022016-01 9-May-2017; 03152016-03 21-Sep-16; 08262016-01 2-Mar-2017; 11042016-01 10-May-2017; 03282016-02 29-Sep-16; 09022016-01 8-Mar-2017; 11092016-01 15-May-2017; 03292016-01 31-Sep-16; 09092016-01 14-Mar-2017; 11142016-02 16-May-2017; 09122016-02; 15-Mar-2017;  11152016-02 18-May-2017; 04012016-01 06-Oct-16; 09192016-01 19-Mar-2017; 11162016-01 21-May-2017; 09192016-02 21-Mar-2017; 11282016-02 30-May-2017; 04062016-01 12-Oct-16; 11302016-02 02-Jun-17; 12052016-01 08-Jun-17; 04112016-04 14-Oct-16; 12052016-03 06-Jun-17; 04142016-01 20-Oct-16; 12072016-01 09-Jun-17; 04202016-02 21-Oct-16; 12092016-01 13-Jun-17; 05022016-02 05-Nov-16; 12132016-01 19-Jun-17; 05092016-01 11-Nov-16; 12152016-02 20-Jun-17; 05162016-02 18-Nov-16; 01042017-01 04-Jul-17; 05232016-01 25-Nov-16; 01052017-03 19-Jul-17; 05272016-01 02-Dec-16; 01052017-04 19-Jul-17; 06022016-02 09-Dec-16; 01092017-01 12-Jul-17; 06102016-01 14-Dec-16; 01102017-01 18-Jul-17; 06152016-01 21-Dec-16; 01132017-01 18-Jul-17; 06172016-01 23-Dec-16; 01132017-03 17-Jul-17; 07282016-02 02-Feb-17.
510(k) numbers
["K131523","K152329","K161202"]
PMA numbers
not on file
Product code
GEH
Firm FEI
3008780134
Firm city
Redwood City
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register