FDA Medical Device Recalls (openFDA)

Siemens Healthcare Diagnostics, Inc: ADVIA Centaur CA 125 II Assay packaged in 100 and 500 test kits as follows: Product code: 09427226 (128532) 100 test kit Product code: 01678114 (128533) 500 Test kit

Recall number
Z-1477-2011
Recalling firm
Siemens Healthcare Diagnostics, Inc
Product
ADVIA Centaur CA 125 II Assay packaged in 100 and 500 test kits as follows: Product code: 09427226 (128532) 100 test kit Product code: 01678114 (128533) 500 Test kit
Status
Terminated
Initiated
2010-12-14
Posted
2011-03-02
Terminated
2013-04-12
Reason
Positive bias (40-55%) with patient results greater than the upper limit of normal and less than the upper limit of normal
Root cause
Process control
Action
Siemens Healthcare Diagnostics on 12 /14/10 issued Urgent Device Recall notice via Federal Express to accounts. This notice informs the customer of the recall and the actions that need to be taken. The Urgent Device Recall notice directs customers to immediately discontinue the use of ADVIA Centaur CA 125 II assay kits that contain CA 125 II reagent lot 144. It also recommends that patient results obtained with this reagent lot be reviewed. A confirmation fax-back form is included to ensure that all customers worldwide have been notified and have complied with the mandatory action. Technical questions, please contact the Technical Solutions Center at 1-877-229-3711
Quantity
2960 units
Distribution
Worldwide distribution, including USA, Canada, Argentina, Austria, Belgium, China, Denmark, France, Germany, Great Britain, Greece, India, Italy, Japan, Korea, Lavitia, Lithuania, Malyasia, Mexico, Netherlands, Norway, Peru, Philippines, Poland, Portugal, Romania, Saudi Arabia, Switzerland, Serbia, Singapore, Slovakia, South Africa, Spain, Sweden, Czech Republic, Taiwan, Turkey, Uganda, and Venezuela
Lot, serial or model codes
Lot Code: 144 Affected kit lot numbers: 62431144, 63124144, 63519144, 63876144, 62450144 and 63822144 Expiry Date: 15 September 2011
510(k) numbers
["K020828"]
PMA numbers
not on file
Product code
LTK
Firm FEI
1219913
Firm city
East Walpole
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register