FDA Medical Device Recalls (openFDA)
Siemens Healthcare Diagnostics, Inc: ADVIA Centaur CA 125 II Assay packaged in 100 and 500 test kits as follows: Product code: 09427226 (128532) 100 test kit Product code: 01678114 (128533) 500 Test kit
- Recall number
- Z-1477-2011
- Recalling firm
- Siemens Healthcare Diagnostics, Inc
- Product
- ADVIA Centaur CA 125 II Assay packaged in 100 and 500 test kits as follows: Product code: 09427226 (128532) 100 test kit Product code: 01678114 (128533) 500 Test kit
- Status
- Terminated
- Initiated
- 2010-12-14
- Posted
- 2011-03-02
- Terminated
- 2013-04-12
- Reason
- Positive bias (40-55%) with patient results greater than the upper limit of normal and less than the upper limit of normal
- Root cause
- Process control
- Action
- Siemens Healthcare Diagnostics on 12 /14/10 issued Urgent Device Recall notice via Federal Express to accounts. This notice informs the customer of the recall and the actions that need to be taken. The Urgent Device Recall notice directs customers to immediately discontinue the use of ADVIA Centaur CA 125 II assay kits that contain CA 125 II reagent lot 144. It also recommends that patient results obtained with this reagent lot be reviewed. A confirmation fax-back form is included to ensure that all customers worldwide have been notified and have complied with the mandatory action. Technical questions, please contact the Technical Solutions Center at 1-877-229-3711
- Quantity
- 2960 units
- Distribution
- Worldwide distribution, including USA, Canada, Argentina, Austria, Belgium, China, Denmark, France, Germany, Great Britain, Greece, India, Italy, Japan, Korea, Lavitia, Lithuania, Malyasia, Mexico, Netherlands, Norway, Peru, Philippines, Poland, Portugal, Romania, Saudi Arabia, Switzerland, Serbia, Singapore, Slovakia, South Africa, Spain, Sweden, Czech Republic, Taiwan, Turkey, Uganda, and Venezuela
- Lot, serial or model codes
- Lot Code: 144 Affected kit lot numbers: 62431144, 63124144, 63519144, 63876144, 62450144 and 63822144 Expiry Date: 15 September 2011
- 510(k) numbers
- ["K020828"]
- PMA numbers
- not on file
- Product code
- LTK
- Firm FEI
- 1219913
- Firm city
- East Walpole
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28