Skip to the content

FDA Medical Device Recalls (openFDA)

Steris Corporation: Steris Lighting and Visualization Systems, Sterile Light handle Cover LB44 -Sterilizable consumables (disposable, single use) used to drape the lighthandle of STERIS surgical lighting systems for the purpose of maintaining a sterile surgical environment. REF: LB44

Recall number
Z-1476-2023
Recalling firm
Steris Corporation
Product
Steris Lighting and Visualization Systems, Sterile Light handle Cover LB44 -Sterilizable consumables (disposable, single use) used to drape the lighthandle of STERIS surgical lighting systems for the purpose of maintaining a sterile surgical environment. REF: LB44
Status
Open, Classified
Initiated
2023-03-17
Posted
2023-04-28
Terminated
not on file
Reason
Light handle covers may separate from the light handle (detach and fall off) during use, this could result in a procedure delay and/or impact the sterile field which could result in an injury to the patient
Root cause
Under Investigation by firm
Action
Steris issued Urgent Medical Device Recall Letter on 3/17/23 via FedEx. Letter states reason for recall, health risk and action to take: 1. Please immediately inspect your on-hand inventory for product affected by this recall. For the full list of lots affected by this recall, please reference Attachment A to this letter. 2. Please complete the Medical Device Recall Response Form included with this letter and destroy any remaining product in your inventory. If product was transferred to any other entities within your organization, please forward this recall notice accordingly. STERIS will coordinate shipment of replacement product upon receipt of the completed Recall Response Form. 3. Return the completed Recall Response Form via email to: Regulatory_Compliance@STERIS.com or via fax to (440) 392-8963. Users are reminded to report any adverse events via FDA s MedWatch Database should an event occur. If you have questions regarding this recall, please contact STERIS Customer Service at 1-800-548-4873, or email the RMA Team at HealthcareRMA@STERIS.com.
Quantity
263,280 covers
Distribution
Worldwide - US Nationwide distribution including Puerto Rico and the countries of Canada, Australia.
Lot, serial or model codes
GTIN number: 00724995192570 Lot Numbers: 9943616 / 9943616A 9943617 / 9943617A 9943619 / 9943619A 9943620 / 9943620A 9959650 / 9959650A 9959651 / 9959651A
510(k) numbers
not on file
PMA numbers
not on file
Product code
FTA
Firm FEI
1527821
Firm city
Mentor
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register