Skip to the content

FDA Medical Device Recalls (openFDA)

Stryker Corporation: Nasopore 4cm fragmentable nasal dressing, Firm Catalog Number 5400-020-004 and Forte Catalog Number 5400-020-004ITL

Recall number
Z-1474-2025
Recalling firm
Stryker Corporation
Product
Nasopore 4cm fragmentable nasal dressing, Firm Catalog Number 5400-020-004 and Forte Catalog Number 5400-020-004ITL
Status
Open, Classified
Initiated
2025-02-25
Posted
2025-03-28
Terminated
not on file
Reason
There is a potential for blister seals on the product to present a bubble on the seal area of the packaging. This seal acts as a sterile barrier and, in this event, indicates that the sterility seal has been breached.
Root cause
Under Investigation by firm
Action
An Urgent Medical Device Recall notification letter dated 2/24/25 was sent to customers. Actions to be taken 1. Immediately review your inventory to locate and quarantine any affected products at your facility. 2. Return the enclosed Business Reply Form (BRF), even if the affected product is no longer in inventory, to the email Instruments.Recalls@stryker.com. 3. Upon receipt of the completed BRF, Stryker will provide a shipping label to return recalled product on-hand. 4. A replacement will be provided upon receipt of the recalled product. 5. Maintain awareness of this communication internally and inform Stryker if any of the subject devices have been distributed to other organizations. If so, provide contact details so Stryker can inform the recipients accordingly. Instruments.recalls@stryker.com. RESPONSE IS REQUIRED.
Quantity
not on file
Distribution
US Nationwide distribution.
Lot, serial or model codes
5400-020-004 UDI-DI: 07613154340811, Lot Numbers: 2022100314 2023013132 2023042514 2023081512 2023110114 2024011514 2024032512 2024082612 2022102514 2023021012 2023050216 2023082811 2023110614 2024011814 2024042914 2023011332 2023021512 2023060214 2023090514 2023111616 2024021213 2024052812 2023012632 2023030814 2023062214 2023091813 2023112112 2024030412 2024073012 ; 5400-020-004ITL UDI-DI: 07613327359459, Lot Numbers: 2022102012 2023011714 2023011714 2023051614 2023081812 2023110914 2024012614 2024030716 2022111514 2023012014 2023012014 2023052314 2023083116 2023112414 2024020714 2024031214 2023010914 2023020214 2023020214 2023062733 2023100914 2023112912 2024021514 2024031814 2023011214 2023051112 2023051112 2023080112 2023101214 2024010912 2024022014 2024032816 2024042314 2024051314 2024052212 2024060512 2024061114
510(k) numbers
not on file
PMA numbers
not on file
Product code
NHB
Firm FEI
3015967359
Firm city
Portage
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register