Skip to the content

FDA Medical Device Recalls (openFDA)

Stryker Corporation: SafeLight Fiber Optic Cable: Clear, 5MM X 10FT 3.05M, REF: 0233-050-100; Gray, 5MM X 10FT 3.05M, REF: 0233-050-200

Recall number
Z-1474-2023
Recalling firm
Stryker Corporation
Product
SafeLight Fiber Optic Cable: Clear, 5MM X 10FT 3.05M, REF: 0233-050-100; Gray, 5MM X 10FT 3.05M, REF: 0233-050-200
Status
Open, Classified
Initiated
2023-03-02
Posted
2023-04-28
Terminated
not on file
Reason
Fiber optic cables assembled with not enough epoxy on the proximal end, creating light output issues, which could cause a clinical delay or procedure being converted to open.
Root cause
Employee error
Action
On 3/2/23 recall notices were mailed to customers who were asked to do the following: 1. Inform individuals within your organization who need to be aware of this action. 2. Identify, remove, and quarantine affected devices. 3. If affected product is found, call customer service at 1-800-624-4422 (Option 3) or email endocustomersupport@stryker.com to arrange for product return. 4. Complete and return the Business Reply Form Customers with questions can contact EndoRecall@stryker.com
Quantity
5,228
Distribution
Worldwide - US Nationwide distribution including in the states of PA, MS, IL, OH, CA, TX, MO, IN, KS, OK, MT, MA, NE, AZ, NV, NJ, IA, WV, MD, KY, FL, NM, NY, CO, WA, NC, VA, LA, ND, WY, HI SC, UT, OR, SD, AR, TN, AK, CT, GA, ID, VT, NH, MN, WI, DE, RI, AL, MI and the countries of NL, PE, AT, CA, TW, PH, SG, KR, CN.
Lot, serial or model codes
Ref/UDI-DI/Lot: 0233-050-100/07613327051650/159190, 159191, 159192, 159194, 159195, 159196, 159197, 159198, 159199, 159200, 159201, 159202, 159203, 159204, 159205, 159206, 159207, 159208, 159209, 159210, 159211, 159212, 159213, 159214, 159215, 159216, 159217, 159218, 164764, 164765, 164766, 164767, 164768, 164769, 164770, 164771, 164772, 164773, 164774, 164775, 164776, 164777, 164778, 164779, 164780, 164781, 164782, 164783, 164784, 164785, 164786, 164787, 164788, 164789, 164790, 164791, 164792, 164793, 164794, 164795, 164796, 164797, 164798, 164799, 164800, 164801, 164802, 164803, 164804, 164805, 164806, 164807, 164808, 164809, 164810, 164811, 164812, 164813, 164814, 164815, 164816, 164817, 164818, 164819, 164820, 164821, 164822, 168940, 168941, 168942, 168943, 168944, 168945, 168946, 168947, 168948, 168949, 168950, 168951, 168952, 168953, 168954, 168955, 168956, 168957, 168958, 168959, 168960, 168961, 168962, 168963, 168964, 168965, 168966, 168967, 168968, 168969, 168970, 168971, 168972, 168973, 168974, 168975, 168976, 168977, 168978, 168979, 168980, 168981, 168987; 0233-050-200/07613327051643/158865, 158868, 158869, 158872, 158874, 158875, 158878, 158884, 165446, 165448, 165451, 165457, 165458, 165459, 165461, 165462, 165464, 168909, 168910, 168911, 168912, 168913, 168914, 168915, 168916, 168917, 168918, 168919, 168920, 168921, 168923, 168924, 168925, 168926, 168927, 168928
510(k) numbers
["K082813"]
PMA numbers
not on file
Product code
FCW
Firm FEI
2936485
Firm city
San Jose
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register