Skip to the content

FDA Medical Device Recalls (openFDA)

Visicu, Inc.: Philips eCareManager System, Model 865325, software versions 3.7, 3.7.1 and 3.8. Product Usage: The Philips eCareManager System software is intended for use in data collection, storage, and clinical information management with independent bedside devices and ancillary systems that are connected either directly or through networks. The eCareManager system provides patient information and surveillance of monitored patients at the point of care location and at a remote, supplementary care location through wide area networking technology and dedicated phone lines and is solely intended for use in a hospital environment. It is not intended to be used in a home environment.

Recall number
Z-1474-2012
Recalling firm
Visicu, Inc.
Product
Philips eCareManager System, Model 865325, software versions 3.7, 3.7.1 and 3.8. Product Usage: The Philips eCareManager System software is intended for use in data collection, storage, and clinical information management with independent bedside devices and ancillary systems that are connected either directly or through networks. The eCareManager system provides patient information and surveillance of monitored patients at the point of care location and at a remote, supplementary care location through wide area networking technology and dedicated phone lines and is solely intended for use in a hospital environment. It is not intended to be used in a home environment.
Status
Terminated
Initiated
2012-02-22
Posted
2012-04-27
Terminated
2012-07-06
Reason
Three versions of software for the eCareManager failed to display anticoagulants on the medications screen. The problem arose due to an unexpected change in the National Drug Data File.
Root cause
Software change control
Action
Philips notified all affected customers via a preliminary email message advising them of the issue. An URGENT - Field Safety Notice follow-up letter was mailed to all customers on February 22, 2012. This letter identified the affected product, described the problem and circumstances in which it can occur; recommend actions the customer should take to prevent risks for patients, and list the actions planned by Philips to correct the problem. For questions contact Philips eICU Customer Support 877-374-2872.
Quantity
12
Distribution
Nationwide Distribution (USA) including states of MA, MO, WI, AR, AZ, PA, SD, VA, and IN.
Lot, serial or model codes
v3.7, v3.7.1, v3.8
510(k) numbers
["K001972","K012171"]
PMA numbers
not on file
Product code
MSX
Firm FEI
3003063823
Firm city
Baltimore
Firm state
MD
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register