Skip to the content

FDA Medical Device Recalls (openFDA)

Ortho-Clinical Diagnostics: VITROS CRP Slides: (1) Cat # 192 6740, 250 slides per box, and (2) Cat # 809 7990, 90 slides per box. Firm on the label: Ortho-Clinical Diagnostics, Inc., Rochester, NY 14626. The CRP Slides are for use on the Vitros 250/250AT and/or Vitros 950/950AT Clinical Chemistry Systems. For in vitro diagnostic use. Vitros CRP slides quantitatively measure C-reactive protein concentration in serum and plasma.

Recall number
Z-1473-04
Recalling firm
Ortho-Clinical Diagnostics
Product
VITROS CRP Slides: (1) Cat # 192 6740, 250 slides per box, and (2) Cat # 809 7990, 90 slides per box. Firm on the label: Ortho-Clinical Diagnostics, Inc., Rochester, NY 14626. The CRP Slides are for use on the Vitros 250/250AT and/or Vitros 950/950AT Clinical Chemistry Systems. For in vitro diagnostic use. Vitros CRP slides quantitatively measure C-reactive protein concentration in serum and plasma.
Status
Terminated
Initiated
2004-07-28
Posted
2004-09-18
Terminated
2005-06-09
Reason
The Immuno-rate (IR) wash detection algorithm may result in inappropriate wash detections.
Root cause
Other
Action
Letters dated 7/28/2004. Distributors instructed to destroy product. Users instructed to use revised IR wash tolerance values (provided with letter). A Calibration Diskette, Data Release Version 5384 will also be provided to end users.
Quantity
Cat 192 6740 - 5,912 cartons x 250 slides; Cat #809 7990 - 10,344 x 90 slides
Distribution
DOMESTIC distribution (nationwide): Product shipped to 6 distributors, 140 hospitals, and 6 U.S. government accounts. FOREIGN distribution: Product shipped to 22 Ortho-Clinical/J&J affiliates in Argentina, Australia, Brazil, Canada, Chile, China, Colombia, Peru, Ecuador, India, Japan, Mexico, Puerto Rico, Panama, Singapore, Venezuela, England, France, Germany, Italy and Spain. The foreign affiliates further distributed product.
Lot, serial or model codes
Coatings 401 thru 409 (Generations 92, 93, 94, 99, 01, 02, 03, 04, and 05)
510(k) numbers
["K030626"]
PMA numbers
not on file
Product code
DCK
Firm FEI
1000136573
Firm city
Rochester
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register