Skip to the content

FDA Medical Device Recalls (openFDA)

Smiths Medical ASD Inc.: Left Heart Kit, Model Number M20268. for blood pressure monitoring.

Recall number
Z-1472-2021
Recalling firm
Smiths Medical ASD Inc.
Product
Left Heart Kit, Model Number M20268. for blood pressure monitoring.
Status
Terminated
Initiated
2021-03-31
Posted
2021-04-21
Terminated
2023-03-20
Reason
Specific lots of Medex High Pressure Adaptors, Rotators, Manifolds, Stopcocks and Tubing and Medex LogiCal Kits may have been manufactured with insufficient or incomplete welds. These are components for various cardiac kits including pressure monitoring devices and interventional imaging devices.
Root cause
Process control
Action
The firm initiated the action by email on 03/31/2021. The Recall Notice instructs consignees to identify, and quarantine affected product in their possession. A list of affected product has been included on the Recall Notice for consignees to identify affected product. The consignees have been instructed to complete the response form, and return it to fieldactions@smiths-medical.com within 10 days of receipt. After the completed Recall Response Form has been submitted to fieldactions@smiths-medical.com, the consignee will be contacted to arrange return of any affected product in their possession. Distributors have been instructed that if they have distributed potentially affected product to their customers to immediately notify them of the Recall.
Quantity
600 devices
Distribution
Distribution US nationwide including Puerto Rico, Austria, Canada, Chile, Ecuador, Hong Kong, Mexico, Slovenia, Brazil, South Africa, Colombia, and Japan.
Lot, serial or model codes
Lot Numbers: 4002486, 4019146, 4036647
510(k) numbers
not on file
PMA numbers
not on file
Product code
NEP
Firm FEI
3012307300
Firm city
Minneapolis
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register