Nova Biomedical Corporation: Varta Easypack XL Lithium Polymer Batteries (P/N 41772 and 42215) Used with: Nova Biomedical StatStrip Glucose Hospital Meter (P/N 42224) and Nova Biomedical StatSensor Creatinine Hospital Meter and EZ CHEM Creatinine Meter (P/N 9600). Statstrip Meter an In vitro diagnostic for use by health care professionals in the quantitative determination of glucose (GLU) in whole blood. EZ Chem In vitro diagnostic for the quantitative measurement of creatinine in capillary, venous or arterial blood.

Recall number
Z-1472-2008
Recalling firm
Nova Biomedical Corporation
Product
Varta Easypack XL Lithium Polymer Batteries (P/N 41772 and 42215) Used with: Nova Biomedical StatStrip Glucose Hospital Meter (P/N 42224) and Nova Biomedical StatSensor Creatinine Hospital Meter and EZ CHEM Creatinine Meter (P/N 9600). Statstrip Meter an In vitro diagnostic for use by health care professionals in the quantitative determination of glucose (GLU) in whole blood. EZ Chem In vitro diagnostic for the quantitative measurement of creatinine in capillary, venous or arterial blood.
Status
Terminated
Initiated
2008-02-21
Posted
2008-09-23
Terminated
2008-10-28
Reason
Lithium batteries may fail if the glucose or creatinine meters have been dropped.
Root cause
Component design/selection
Action
Nova Biomedical notified consignees by telephone and followed by the Recall Notification letter for the Vata Easy Pack XL Lithium Polymer Battery, dated 2/2108 and mailed 2/26/08. Users are notified of the problem with the battery in the Nova Stat Strip Blood Glucose Hospital Meter System and is replacing batteries by on site visit. E-Z-EM , Inc. notified their customers by letter dated 3/17/08. Contact Nova Biomedical Corporation at 1-800-545-6682 for assistance.
Quantity
4432 batteries
Distribution
Worldwide Distribution --- including USA and countries of Canada, Taiwan, Hungary, Malaysia, Australia, Russia, Thailand, New Zealand, Austria, France, Germany, UK, China, South Africa, Croatian, Peru, and Czech Republic.
Lot, serial or model codes
Batteries coded: XX07 and earlier The Nova Biomedical products that contain the lithium batteries include the following part numbers: Part # 41772 - Battery Assy,  Part # 42215 - Battery Replacement pack,  Part # 42224 - Kit Meter,  Part # 43735 - Kit Meter EZ CHEM,  Part # 43773 - Kit Meter,  Part # 43811 - Battery Replacement Pack EZ CHEM,  Part # 44013 - Kit Meter,  Part # 44014 - Kit Meter and Part # 44251 - Kit Meter.
510(k) numbers
["K060345","K070068"]
PMA numbers
not on file
Product code
CGA
Firm FEI
1219029
Firm city
Waltham
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register