Abiomed, Inc.: Impella RP. Product Code: 0046-0011.

Recall number
Z-1471-2026
Recalling firm
Abiomed, Inc.
Product
Impella RP. Product Code: 0046-0011.
Status
Open, Classified
Initiated
2026-01-27
Posted
2026-02-27
Terminated
not on file
Reason
Differential pressure (dP) sensor in Impella RP devices may malfunction, causing sensor values to drift.
Root cause
Under Investigation by firm
Action
Abiomed notified consignees on about 01/27/2026 via UPS delivery. Additionally, the Customer Letter was posted on Heartrecovery.com: https://www.heartrecovery.com/en-us/resources/downloads/field-safety-notification-impella-rp-differential-pressure-sensor. The notifications informed consignees were instructed clinicians that affected units can continue use and they should: 1. Continue to monitor patient hemodynamics with approved diagnostic devices and continue to verify Impella positioning with imaging before clinical interventions. 2. Refer to the P-level flow rates listed in the Instructions for Use (IFUs), as shown in Figure 1, and rely on these values rather than the AIC display. 3. Trend pump metrics to monitor pump performance and to identify sudden change, noting that abrupt changes in the displayed flow rate need to be assessed both against the flow rate listed in the IFU and across temporal trends. Consignees were instructed that product is NOT being removed from the field and does not need to be returned, patients should be assessed with approved diagnostic devices before clinical interventions, complete and return the provided business reply form (BRF) to Abiomed4437@sedgwick, forward the notification to anyone in the facility that needs to be informed (i.e., those who manage, transport, store, stock, or use the subject products) and if any of the subject products have been forwarded to another facility, contact that facility and provide them with this notice.
Quantity
179 units
Distribution
Worldwide distribution - US Nationwide and the countries of Austria, Canada, Croatia, Czechia, France, Germany, Hong Kong, Italy, Poland, Serbia, Switzerland.
Lot, serial or model codes
Product Code: 0046-0011. UDI-DI: 04260113630273. Distributed outside the US.
510(k) numbers
not on file
PMA numbers
["P170011"]
Product code
PYX
Firm FEI
1220648
Firm city
Danvers
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register