FDA Medical Device Recalls (openFDA)
Heartware, Inc.: Instructions for Use and Patient Manual for HeartWare HVAD System. Accompanies these devices: Heartware HVAD pump implant kits Models 1100, 1101, 1102, 1103, 1104, 1104JP, 1205, and MCS1705PU. Heartware HVAD controller kits Models 1403US, 1407AU, 1407CA, 1407CH, 1407DE, 1407GB, 1407IL, 1407IN, 1407IT, 1407JP, 1407KR, 1420, and 1420JP. Heartware HVAD controller AC adapter Models 1430AR, 1430AU, 1430CA, 1430CH, 1430DE, 1430GB, 1430IL, 1430IN, 1430IT, 1430JP, and 1430US. Heartware HVAD controller DC Adapter Model 1440. Heartware HVAD data cable - monitor Model 1575. Heartware HVAD battery pack Models 1650, 1650CA-CLIN, and 1650DE. Heartware HVAD patient pack Model 1475. Heartware HVAD waist pack Models 2050, 2050IL, and 2050US. Heartware HVAD shoulder pack Models 2060, 2060IL, and 2060US.
- Recall number
- Z-1469-2021
- Recalling firm
- Heartware, Inc.
- Product
- Instructions for Use and Patient Manual for HeartWare HVAD System. Accompanies these devices: Heartware HVAD pump implant kits Models 1100, 1101, 1102, 1103, 1104, 1104JP, 1205, and MCS1705PU. Heartware HVAD controller kits Models 1403US, 1407AU, 1407CA, 1407CH, 1407DE, 1407GB, 1407IL, 1407IN, 1407IT, 1407JP, 1407KR, 1420, and 1420JP. Heartware HVAD controller AC adapter Models 1430AR, 1430AU, 1430CA, 1430CH, 1430DE, 1430GB, 1430IL, 1430IN, 1430IT, 1430JP, and 1430US. Heartware HVAD controller DC Adapter Model 1440. Heartware HVAD data cable - monitor Model 1575. Heartware HVAD battery pack Models 1650, 1650CA-CLIN, and 1650DE. Heartware HVAD patient pack Model 1475. Heartware HVAD waist pack Models 2050, 2050IL, and 2050US. Heartware HVAD shoulder pack Models 2060, 2060IL, and 2060US.
- Status
- Open, Classified
- Initiated
- 2021-02-26
- Posted
- not on file
- Terminated
- not on file
- Reason
- IFU and Patient Manual updated to describe Expected Alarm Duration. The Instructions for Use (IFU) is being updated to advise clinicians that alarms that resolve after a very short period may mean there is an intermittent problem and the cause of the alarm can be evaluated by analysis of the controller log files. The Patient Manual (PM) is being updated to instruct patients to contact their clinicians if they cannot identify an alarm before it is cleared as it might mean there is an intermittent problem. NOTE: this update is included in the February 2021 Urgent Medical Device Notice also covered in recalls 87456 and 87604
- Root cause
- Device Design
- Action
- Beginning 26-Feb-2021, an Urgent Medical Device Recall notification was disseminated to consignees who Medtronic records indicate have received affected product, UPS 2-day delivery. The notices provided updated information for the instructions for use and patient manual. Note: this notice covers issues in three separate recalls in FDA's database.
- Quantity
- 23123 (11292 US)
- Distribution
- Worldwide distribution, including: US nationwide, Italy, Great Britain, Germany, Japan
- Lot, serial or model codes
- All serial numbers as of February 26, 2021. Model, GTIN: 1100, No GTIN Available; 1101, No GTIN Available; 1102, No GTIN Available; 1103, 00888707000017; 1104, No GTIN Available; 1104JP, 00888707000697; 1205, No GTIN Available; MCS1705PU, 00888707005364; 1403US, 00888707000475; 1407AU, 00888707001663; 1407CA, 00888707002851; 1407CH, 00888707001694; 1407DE, 00888707001700; 1407GB, 00888707001717; 1407IL, 00888707001724; 1407IN, 00888707001731; 1407IT, 00888707001748; 1407JP, 00888707001755; 1407KR, 00888707001762; 1420, 00888707000420; 1420JP, 00888707000437; 1430AR, 00888707000826; 1430AU, 00888707000833; 1430CA, 00888707000857; 1430CH, 00888707000871; 1430DE, 00888707000864; 1430GB, 00888707000888; 1430IL, 00888707000901; 1430IN, 00888707000505; 1430IT, 00888707000895; 1430JP, 00888707000918; 1430US, 00888707000307; 1440, 00888707001885; 1575, 00888707008051; 1650, 00888707000376; 1650CA-CLIN, 00888707001588; 1650DE, 00888707001373; 1475, 00888707008037; 2050, 00888707008082; 2050IL, 00888707001410; 2050US, 00888707000048; 2060, 00888707008099; 2060IL, 00888707001427; 2060US, 00888707000055;
- 510(k) numbers
- not on file
- PMA numbers
- ["P100047"]
- Product code
- DSQ
- Firm FEI
- 3007042319
- Firm city
- Miami Lakes
- Firm state
- FL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28