The Binding Site Group, Ltd.: VaccZyme Human Anti-Haemophilus influenza type b Enzyme Immunoassay Kit For in vitro diagnostic use, Product Code: MK016 - Product Usage: This assay is designed for the in vitro measurement of specific IgG antibodies against Haemophilus influenzae type b (Hib) capsular polysaccharide, present in human serum. Sufficient materials are supplied to allow a maximum of 41 samples to be tested in duplicate, with a calibration curve and two controls.

Recall number
Z-1469-2020
Recalling firm
The Binding Site Group, Ltd.
Product
VaccZyme Human Anti-Haemophilus influenza type b Enzyme Immunoassay Kit For in vitro diagnostic use, Product Code: MK016 - Product Usage: This assay is designed for the in vitro measurement of specific IgG antibodies against Haemophilus influenzae type b (Hib) capsular polysaccharide, present in human serum. Sufficient materials are supplied to allow a maximum of 41 samples to be tested in duplicate, with a calibration curve and two controls.
Status
Terminated
Initiated
2018-08-12
Posted
not on file
Terminated
2022-01-20
Reason
A deterioration of performance was identified with influenza type B Enzyme Immunoassay Kits with findings of a positive bias with kit controls. Patient sample results may be affected with falsely elevated results incorrectly indicating a protective level of anti-Hib antibody in the assessment of immunodeficiency.
Root cause
Under Investigation by firm
Action
On XXX, an Urgent Field Safety Notice was sent via email to customers. Customers are instructed to discontinue using this product immediately; dispose of any remaining product at your facility following local regulations; discuss this information with your medical director so that the potential impact of the issue is considered on a local level; and to return the completed and signed E-Back form TSWS18 to Technical.Support@bindingsite.com or to your local firm representative within one week of receiving this notification. Customers are also requested to include a witness signature for the kit disposal as this is required on the E-Back form. Customers with additional questions were encouraged to email Technical.support@bindingsite.com.
Quantity
462 kits
Distribution
Worldwide distribution - US Nationwide distribution in the states of M.A., C.A., MO., NC, MN, VA, WA, and IL and country of Canada.
Lot, serial or model codes
UDI: 05051700006767. Lot numbers 413170 and 435314.
510(k) numbers
not on file
PMA numbers
not on file
Product code
GRP
Firm FEI
3002808340
Firm city
Birmingham
Firm state
not on file
Firm country
United Kingdom

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register