Skip to the content

FDA Medical Device Recalls (openFDA)

Churchill Medical Systems, Inc.: 4" Microbore Double Lumen Set with Bionector, Product Code: BN-208 Accessory device used to administer medical fluids

Recall number
Z-1469-2015
Recalling firm
Churchill Medical Systems, Inc.
Product
4" Microbore Double Lumen Set with Bionector, Product Code: BN-208 Accessory device used to administer medical fluids
Status
Terminated
Initiated
2015-01-15
Posted
2015-04-21
Terminated
2016-02-25
Reason
Potential for leaking caused by insufficient bond between needleless device and female luer.
Root cause
Process design
Action
Vygon sent an Urgent Medical Device Recall letter dated January 15, 2015, to all affected customers via email or courier service. The letter instructs consignees to check stock for products included in the scope of the recall, count and document the affected inventory on the "Recall Acknowledgement and Inventory Return Form" and fax or email the form to Vygon. Vygon customer service will then contact the customer to issue a Return Merchandise Authorization Number so the inventory can be replaced at no cost. Distributors were provided with instructions on how to handle product that was further distributed by them. If after 30-days, product is not returned from each consignee, Vygon will follow-up with the consignee to obtain a status of the product return. Customers with questions were instructed to contact Vygon Customer Service at 1-800-473-5414, or by e-mail at customerservice@vygonus.com.
Quantity
2350
Distribution
Nationwide Distribution including TX, OH, NY, CA, UT, PA, MA, NY, VA, and IL.
Lot, serial or model codes
1303123 1305003 1305008 1305049 1306097
510(k) numbers
["K021395","K070705","K963981"]
PMA numbers
not on file
Product code
FPA
Firm FEI
1000120555
Firm city
Dover
Firm state
NH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register