Brainlab AG: Brainlab, Spine & Trauma 3D 2.0, Navigation Software. An intraoperative image-guided localization system to enable minimally invasive surgery. Catalog Number: 22264

Recall number
Z-1469-2014
Recalling firm
Brainlab AG
Product
Brainlab, Spine & Trauma 3D 2.0, Navigation Software. An intraoperative image-guided localization system to enable minimally invasive surgery. Catalog Number: 22264
Status
Terminated
Initiated
2013-11-18
Posted
2014-04-17
Terminated
2017-07-17
Reason
The "fluoro match registration" function of the Navigation Software Spine & Trauma 3D 2.0 allows the user to intraoperatively match e.g., CT data sets to the current patient anatomy visible on fluoroscopic images, used by the navigation software during spinal surgeries to display the position and orientation of instruments. For this registration function in combination with a digitally integrated
Root cause
Software design
Action
Brainlab sent a FIELD SAFETY NOTICE / PRODUCT NOTIFICATION letter dated November 18, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The letter included instructions for the modification of C-arm configurations: 1) If possible, Use analog image transfer between C-arm and Brainlab Navigation System and to consult with the local Brainlab Customer Support Representative for this modification, and 2) If possible, adjust the image size of the C-arm to 8 bit and to follow the instructions provided with the C-arm and if required to consult the C-arm manufacturer. For measures when performing a Fluoro match registration, the following instructions are provided to reduce the probability that the algorithm results in an inaccurate match: 1) perform the preoperative CT scans according to the descriptions in the Brainlab scan protocol, 2) set the threshold so that a smooth bone surface is displayed and the quality of the chosen bone threshold influences the accuracy of the match result (refer to section "Bone Threshold" in the Software User Guide"), 3) use the Radiolucent Spine Reference Clamp instead of the Spine Reference X-Clamp, 4) make sure that no additional objects (e.g. wires, retractors, instruments, etc.) are within the view of the C-arm. Additional reminders in the letter include: 1) always verify registration accuracy by holding the pointer or instrument tip to at least three anatomical landmarks and verifying their position in the software, 2) verify that the registration is at the correct level on the patient and data set, and 3) accuracy must be checked on the treated bone surface. Brainlab will provide a software update with the issue solved to affected customers tentatively by January 2014. Customers with questions can contact the Customer Hotline at 800-597-5911 or via e-mail at us.support@brainlab.com.
Quantity
133 units
Distribution
Worldwide Distribution - US Distribution including the states of AZ, CA,GA, IL, IN, LA, MA, MI, NY, NC, OH, PA, TN, TX and WI and the countries of Australia, Austria, Belgium, Brazil, Chile, Czech Republic, France, Germany, Italy, Japan, Malaysia, Netherlands, Norway, Philippines, Poland, Russian Federation, Singapore, South Africa, Spain, Sweden, Switzerland, Thailand, Turkey and United Kingdom.
Lot, serial or model codes
Catalog Number: 22264; Serial Numbers: 000000000038012001 , 000000000060912001 , 000000000189912001 , 000000000224912001 , 000000000235512001 , 000000000282912001 , 000000000313612001 , 000000000319412001 , 000000000351112001 , 000000000356912001 , 000000000367212001 , 000000000433712001 , 000000000482012001 , 000000000502812001 , 000000000645612001 , 000000000731112001 , 000000000837312001 , 000000000898312001 , 000000000956812001 , 000000000998412001 , 000000001012812001 , 000000001105412001 , 000000001231812001 , 000000001353512001 , 000000001355412001 , 000000001433712001 , 000000001473712001 , 000000001493512001 , 000000001493712001 , 000000001566212001 , 000000001579112001 , 000000001635112001 , 000000001758212001 , 000000001770212001 , 000000001789312001 , 000000001868312001 , 000000001895412001 , 000000001909912001 , 000000002036112001 , 000000002309812001 , 000000002439112001 , 000000002468312001 , 000000002503012001 , 000000002573012001 , 000000002682913001 , 000000003000913001 , 000000003053313001 , 000000003066813001 , 000000003071613001 , 000000003087613001 , 000000003109513001 , 000000003109613001 , 000000003230113001 , 000000003246713001 , 000000003304113001 , 000000003331913001 , 000000003379313001 , 000000003453413001 , 000000003453513001 , 000000003622213001 , 000000003671813001 , 000000003744413001 , 000000003758413001 , 000000003819613001 , 000000003824913001 , 000000003866013001 , 000000003937813001 , 000000004094013001 , 000000004147413001 , 000000004356413001 , 000000004404013001 , 000000004489313001 , 000000004489413001 , 000000004505313001 , 000000004821913001 , 000000004882013001 , 000000004905013001 , 000000005396313001 , 000000005407413001 , 000000005505113001 , 000000005506013001 , 000000005531913001 , 000000005552313001 , 000000005578713001 , 000000005659713001 , 000000005702513001 , 000000005788213001 , 000000005817113001 , 000000005904513001 , 000000005904613001 , 000000005904713001 , 000000005904813001 , 000000005946713001 , 000000006046413001 , 000000006046513001 , 000000006096913001 , 000000006136313001 , 000000006233113001 , 000000006296713001 , 000000006426913001 , 000000006427213001 , 000000006621613001 , 000000006621613002 , 000000006656613001 , 000000006792613001 , 000000006833413001 , 000000006839413001 , 000000006998813001 , 000000007000113001 , 000000007122113001 , 000000007172913001 , 000000007405013001 , 000000007588513001 , 000000007637114001 , 000000007679714001 , 000000007698714001 , 000000007784614001 , 000000007859414001 , 000000007906314001 , 000000008020214001 , 000000008031314001 , 000000008184514001 , 000000008193214001 , 000000008199814001 , 000000008214014001 , 000000008259514001 , 000000008383314001 , 000000009718312001 , 000000009720812001 , 000000009781112001 , 000000009802512001 , 000000009901612001 , 000000009983912001
510(k) numbers
["K070106"]
PMA numbers
not on file
Product code
HAW
Firm FEI
3002619595
Firm city
Feldkirchen
Firm state
not on file
Firm country
Germany

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register